72
Teledyne API – Model T400 O
3
Analyzer
083730400A DCN8130
Figure 2-45. Analog Outputs Group Calibration Screen
Figure 2-46. Analog Outputs Manual Calibration Screen (AOUT2 Example)
Table 2-11. Analog Output Voltage/Current Range
For manual calibration adjustments, see Section 2.5.8.1 for voltage and Section 2.5.8.2
for current.
RANGE
1
RANGE SPAN
MINIMUM OUTPUT
MAXIMUM OUTPUT
100mV
0-100 mVDC
-5 mVDC
105 mVDC
1V
0-1 VDC
-0.05 VDC
1.05 VDC
5V
0-5 VDC
-0.25 VDC
5.25 VDC
10V
0-10 VDC
-0.5 VDC
10.5 VDC
Current
2
0-20 mA
0 mA
20 mA
1
Each range is usable from -5% to +5% of the rated span.
2
While these are the physical limits of the current loop modules, typical applications use 2-20 mA or 4-20
mA for the lower and upper limits.