40
Teledyne API – Model T400 O
3
Analyzer
083730400A DCN8130
For the Model T400 photometric ozone analyzer in its basic configuration (i.e. without
the optional internal zero air source or valves), attach the following pneumatic lines:
S
AMPLE
G
AS
S
OURCE
Attach a sample inlet line to the
SAMPLE
inlet port.
•
Sample Gas pressure must equal ambient atmospheric pressure (1.0 psig)
•
In applications where the sample gas is received from a pressurized manifold, a vent must
be placed on the sample gas line. This vent line must be:
•
At least 0.2m long
•
No more than 2m long
•
Vented outside the shelter or immediate area surrounding the instrument
C
ALIBRATION
G
AS
S
OURCES
The source of calibration gas is also attached to the
SAMPLE
inlet, but only when a
calibration operation is actually being performed.
E
XHAUST
O
UTLET
Attach an exhaust line to the analyzer’s EXHAUST outlet fitting. The exhaust line
should be:
•
PTFE tubing; minimum O.D ¼”;
•
A maximum of 10 meters long;
•
Vented outside the analyzer’s enclosure.
L
EAK
C
HECK
Once the appropriate pneumatic connections have been made, check all pneumatic
fittings for leaks using the procedures defined in Section 5.6.3.