083730400A DCN8130
Teledyne API – Model T400 O
3
Analyzer
45
2.3.2.5 PNEUMATIC CONNECTIONS: AMBIENT AIR MONITORING IN REMOTE LOCATIONS
In this application it is often preferred that the Sample gas and the source gas for the O
3
generator and internal zero air be the same chemical composition.
Enclosure Wall
Instrument Chassis
(with IZS installed)
ZERO AIR
DRY AIR
SAMPLE
EXHAUST
SPAN
LOCATION
BEING
MONITORED
Maximum 15 ft. in length
Figure 2-17. Gas Line Connections when the T400 Analyzer is Monitoring a Remote Location
S
AMPLE
G
AS
S
OURCE
Attach a sample inlet line leading from the room being monitored to the
SAMPLE
inlet
fitting.
D
RY
A
IR
S
OURCE
Attach a gas line leading from the room being monitored to the
DRY AIR
inlet port. This
can be a separate line or, as shown above (Figure 2-17) in the same line with a T-fitting.
E
XHAUST
O
UTLET
No outlet line is required for the exhaust port of the instrument.
L
EAK
C
HECK
Once the appropriate pneumatic connections have been made, check all pneumatic
fittings for leaks using the procedures defined in Section 5.6.3.