083730400A DCN8130
Teledyne API – Model T400 O
3
Analyzer
159
Figure 6-15. Typical Jumper Set (JP2) Set Up of Optional Metal Wool Scrubber Heater
6.3.6 P
HOTOMETER
L
AYOUT AND
O
PERATION
The Photometer is the component where the absorption of UV light by ozone is measured
and converted into a voltage. It consists of several sub-assemblies:
•
A mercury-vapor UV lamp. This lamp is coated in a material that optically screens the UV
radiation output to remove the O
3
producing 185nm radiation. Only light at 254nm is
emitted.
•
An AC power supply to supply the current for starting and maintaining the plasma arc of
the mercury vapor lamp.
•
A thermistor and DC heater attached to the UV lamp to maintain the lamp at an optimum
operating temperature.
•
42 cm long quartz absorption tube.
•
A thermistor attached to the quartz tube for measuring sample gas temperature.
•
Gas inlet and outlet mounting blocks that route sample gas into and out of the photometer.
•
The vacuum diode, UV detector that converts UV light to a DC current.
•
A preamplifier assembly, which convert the Detector’s current output into a DC Voltage
then amplifies it to a level readable by the A-to-D converter circuitry of the instrument’s
motherboard.
CAUTION – UV RADIATION RISK
Do not look at the UV lamp while the unit is operating. UV light can
cause eye damage. Always use safety glasses made from UV
blocking material when working with the UV Lamp Assembly.
(Generic plastic glasses are not adequate).
Содержание T400
Страница 2: ......
Страница 101: ...99 Teledyne API Model T400 O3 Analyzer 083730400A DCN8130 This page intentionally left blank ...
Страница 176: ...This page intentionally left blank A 4 083730400A DCN8130 ...
Страница 177: ...Appendix B Wiring Diagram 083730400A DCN8130 ...