083730400A DCN8130
Teledyne API – Model T400 O
3
Analyzer
113
8. Locate the potentiometer used to adjust the O
3
generator UV output.
Adjustment
Pot
O
3
Generator
Reference
Detector
PCA
O
3
Generator
Body
Figure 5-10. Location of O
3
Generator Adjustment Pot
9. Navigate to Utilities>Diagnostics>Analog Inputs>O3 Gen Ref.
10. Using an insulated pot adjustment tool, turn the O
3
Generator Adjustment Pot
(Figure 5-10) until the value of O3 Gen Ref is 2500.0 mV ±100 mV.
11. Finish screwing in the UV lamp holder so that it is snug but not excessively tight.
12. Replace the analyzer’s cover.
13. Perform a leak check (Section 5.6.5).
14. Perform an ozone generator calibration (Section 4.5.3).