112
Teledyne API – Model T400 O
3
Analyzer
083730400A DCN8130
5.6.10
A
DJUSTING OR
R
EPLACING THE
IZS
O
ZONE
G
ENERATOR
O
PTION
’
S
UV
L
AMP
This procedure details the steps for replacing and calibrating the UV lamp of the O
3
Generator that is included in the Internal Zero Span (IZS) option. If adjusting an existing
lamp, skip to Step 8.
CAUTION – Mercury Exposure Risk
The UV lamp contains mercury (Hg), which is considered hazardous
waste. The lamp should be disposed of in accordance with local
regulations regarding waste containing mercury.
1. Power off the analyzer.
2. Remove the cover from the analyzer.
3. Locate the O
3
generator (see Figure 2-3).
Figure 5-9. O
3
Generator UV Lamp
4. Unscrew the UV lamp holder and gently pull out the old lamp.
5. Inspect the o-ring beneath the lamp holder and replace if damaged.
6. Install the new lamp in O
3
generator housing, and partially tighten the UV lamp
holder.
•
The lamp should be free enough to be rotated in the assembly by grasping the
lamp cable.
7. Power on the analyzer and allow it to stabilize for at least 20 minutes.