083730400A DCN8130
Teledyne API – Model T400 O
3
Analyzer
131
5.7.13 T
ROUBLESHOOTING THE
P
HOTOMETER
This section involves checking the measure/reference valve and the photometer UV lamp
power supply.
5.7.13.1 CHECKING THE MEASURE/REFERENCE VALVE
To check the function of the photometer’s measure / reference valve:
1. View the
O3 REF
function in the Dashboard (Setup>Dashboard to configure).
2. Follow the instruction in Section 4.2.1.1 for performing a zero point calibration check.
•
Press
ZERO
and allow the analyzer to stabilize.
3. Note the displayed value.
4. Follow the instruction in Section 4.2.1.2 for performing a span point calibration check.
•
Press
SPAN
and allow the analyzer to stabilize.
5. Note the displayed value for
O3 REF
.
•
If the
O3 REF
value has decreased by more than 2 mV from its value with zero gas,
then there is a "cross-port" leak in the M/R valve or a bad O
3
reference scrubber.
Refer to Section 5.7.15.2 for replacement instructions.
•
Press the
EXIT
button to interrupt the span point calibration process (
DO NOT
PRESS the ENTR button).
5.7.13.2 CHECKING THE PHOTOMETER UV LAMP POWER SUPPLY
WARNING – ELECTRICAL SHOCK HAZARD
Hazardous voltage present - use caution.
CAUTION – UV RADIATION RISK
Do not look at the UV lamp while the unit is operating. UV light can
cause eye damage. Always use safety glasses made from UV
blocking material when working with the UV Lamp Assembly.
(Generic plastic glasses are not adequate).
Although it is not always possible to determine with certainty whether a UV lamp
problem lies with the lamp or with its power supply, the following steps will provide a
reasonable confidence test of the lamp power supply.
1. Unplug the cable connector at P1 on the lamp power supply and confirm that +15VDC is
present between Pins 1 and 2 on the cable connector.
2. If this voltage is incorrect, check the DC test points on the relay PCA as described in
Section 5.7.9.2