102
Teledyne API – Model T400 O
3
Analyzer
083730400A DCN8130
Figure 5-2. Remote Update Page
5.4.2 M
ANUAL
R
ELOAD
/U
PDATE
P
ROCEDURES
To reload or update firmware, first contact Technical Support to obtain the applicable
file(s): [email protected] /+1 800-324-5190.
1. Follow Technical Support’s instructions for copying the firmware files to a
flash drive.
2. Go to the Utilities>USB Utilities>General menu.
Figure 5-3. Manual Update Page (and other utilities)
3. Insert a flash drive into a front panel USB port and wait for the Status field to
indicate that the drive has been detected.
4. In the Update Firmware field, press the Check button for the instrument to
determine whether the firmware on the flash drive is more recent than what is
currently installed. Once it’s been determined that the firmware is new, the
Install button will be enabled; if the firmware version on the flash drive is the
same as or older than the current firmware of the instrument, the Install
button will not be enabled.
5. Press the Install button, and note the messages in the Status field at the
bottom of the page. Use the Cancel button if necessary.
6. When complete, as indicated in the Status field, press the Done button,
which replaces the Cancel button, and remove the flash drive.
7. Power off and restart the instrument to complete the new firmware
installation.