48
Teledyne API – Model T400 O
3
Analyzer
083730400A DCN8130
2.3.3.3 PNEUMATIC FLOW: INTERNAL ZERO/SPAN (IZS) VALVE OPTION
INSTRUMENT CHASSIS
Flow / Pressure
Sensor PCA
SAMPLE
PRESSURE
SENSOR
O
3
FLOW
SENSOR
Sample Gas
Flow Control
EXHAUST
GAS
OUTLET
Particulate
Filter
PUMP
DRY AIR
INLET
O
3
Scrubber
A
B
S
O
RP
TI
O
N
TUB
E
SAMPLE
GAS INLET
SPAN GAS
INLET
ZERO AIR
INLET
Sample/Cal
Valve
Measure/
Reference
Valve
Charcoal
Scrubber
& Filter
O
3
Generator
Figure 2-20. T400 Pneumatic Diagram with Internal Zero/Span (IZS) Option (OPT50G)
Table 2-7. Internal Zero/Span Valve Operating States for Option 50G
Mode
Valve
Condition
SAMPLE
Sample/Cal Valve
Open to SAMPLE inlet
Ozone Generator
OFF
ZERO CAL
Sample/Cal Valve
Open to Ozone Generator
Ozone Generator
OFF
SPAN CAL
Sample/Cal Valve
Open to Ozone Generator
Ozone Generator
ON at intensity level set by user
The state of the Sample/Cal valves can be controlled:
•
Manually via the analyzer’s front panel;
•
By activating the instrument’s AutoCal feature (See Section 4.3);
•
Remotely by using the External Digital I/O Control Inputs (See Section 4.2.2.1)