083730400A DCN8130
Teledyne API – Model T400 O
3
Analyzer
47
2.3.3.2 PNEUMATIC FLOW: ZERO/SPAN (ZS) VALVE OPTION
INSTRUMENT CHASSIS
Flow / Pressure
Sensor PCA
SAMPLE
PRESSURE
SENSOR
O
3
FLOW
SENSOR
Sample Gas
Flow Control
EXHAUST
GAS
OUTLET
Particulate
Filter
PUMP
DRY AIR
INLET
O
3
Scrubber
A
B
S
O
RP
TI
O
N
TUB
E
SAMPLE
GAS INLET
SPAN GAS
INLET
ZERO AIR
INLET
Zero/Span
Valve
Sample/Cal
Valve
Measure/
Reference
Valve
Figure 2-19. T400 Pneumatic Diagram with Zero/Span (ZS) Option (OPT50A)
Table 2-6. Zero/Span Valve Operating States for Option 50A
Mode
Valve
Condition
SAMPLE
Sample/Cal
Open to SAMPLE inlet
Zero/Span
Open to ZERO AIR inlet
ZERO CAL
Sample/Cal
Open to ZERO/SPAN valve
Zero/Span
Open to ZERO AIR inlet
SPAN CAL
Sample/Cal
Open to ZERO/SPAN valve
Zero/Span
Open to SPAN GAS inlet
The state of the Sample/Cal valves can be controlled:
•
Manually via the analyzer’s front panel;
•
By activating the instrument’s AutoCal feature (See Section 4.3);
•
Remotely by using the External Digital I/O Control Inputs (See Section 4.2.2.1)