AFA4000/E
VAV Airflow Controller / p.99
12.
If the sample is accepted, lower the sash by half and measure the face velocity using a
calibrated instrument. Use the
↑
/
↓
buttons to enter the measured face velocity and press
Mute
. The monitor samples the airflow for 5 seconds.
13.
If the calibration is successful, the monitor returns to the
Main Menu
. Select
Run
to go to
normal operating mode and check the velocity reading is accurate and stable.
14.
If the calibration is stable and accurate, press and hold the
Mute
button for 5 seconds until
the
Main Menu
is displayed.
15.
Use the
↑
/
↓
buttons to select
Set Up Monitor
, then
Configure
.
16.
Enter the password (the factory default password is 0-0-0-0) and press
Mute
to continue.
17.
Use the
↑
/
↓
buttons to select
Econ Configure
, then
Manual/Auto
followed by
Mute
.
18.
Select
Auto
and press
Mute
to continue.
19.
Select
Done
and press
Mute
.
20.
Select
Run
to go to normal operating mode.
7.1.1
Possible messages
Message
Reason
Actions
Deviations too high
Airflow sample is unstable.
Follow the instructions to repeat
the sample or quit the
calibration.
Increase higher airflow
sample
Second sample value is too
close to the first sample value.
Close the sash slightly and repeat
the higher sample. The minimum
difference between the samples
that the controller will accept is
0.3 m/sec (60 fpm).
Sensor diff too low
The monitor does not detect
any difference in the sensor
output between the 2 airflow
samples.
Check that the sensor hose is
connected and repeat the
calibration.
7.1.2
Calibration notes
•
Make sure the sensor hose is connected correctly at the side wall and rear of the sensor.
•
Make sure the extraction fan is running and the fume cupboard is balanced before
calibrating the monitor.
Содержание AFA4000/E
Страница 1: ...AFA4000 E VAV Airflow Controller Installation and operating manual Issue 2...
Страница 57: ...AFA4000 E VAV Airflow Controller p 57 Figure 22 AFA4000 E inverter control connections...
Страница 58: ...AFA4000 E VAV Airflow Controller p 58 Figure 23 AFA4000 E damper control connections...
Страница 59: ...AFA4000 E VAV Airflow Controller p 59 Figure 24 AFA4000 E typical connection diagram...
Страница 60: ...AFA4000 E VAV Airflow Controller p 60 Figure 25 AFA4000 E AS typical connection diagram...
Страница 62: ...AFA4000 E VAV Airflow Controller p 62 Figure 27 SM7 airflow sensor dimensions...
Страница 71: ...AFA4000 E VAV Airflow Controller p 71 Figure 34 Side views of a rack and pinion drive typical installation...
Страница 96: ...AFA4000 E VAV Airflow Controller p 96 6 9 Auto Sash System Control Box general wiring diagram...
Страница 158: ...AFA4000 E VAV Airflow Controller p 158...
Страница 182: ...AFA4000 E VAV Airflow Controller p 182 Figure 79 Connection diagram with RS232 485 converter...