AFA4000/E
VAV Airflow Controller / p.14
3.3
Alarms and events
Message
Displayed when
Sash high
The Sash alarm is activated, and the sash is raised above the maximum
safe working opening.
Sash high
alternates on/off with the velocity
reading.
External alarm
The external alarm input is activated (when enabled).
External alarm
alternates on/off with the velocity reading.
Air safe
The airflow is at a safe velocity.
Air Safe
alternates on/off with the
velocity reading.
Air fail
The airflow is less than the low air alarm point.
Air Fail
alternates on/off
with the velocity reading.
High air (alarm)
The airflow is more than the high air alarm point.
High air
alternates
on/off with the velocity reading.
High Air
The airflow is more than the high air cut off value.
High air
is displayed
without the velocity reading.
Setback
The night setback function is activated.
Setback
alternates on/off with
the velocity reading. The display backlight dims.
Standby
Pushbutton 3 Min function is activated.
Standby
alternates on/off with
the velocity reading. The display backlight dims.
Alarm disabled
The alarm disable function is activated.
Alarm disable
alternates on/off
with the velocity reading.
Close sash
The sash is raised and the operator is not present (when activated).
Close
sash
alternates on/off with the velocity reading.
Emergency
The emergency input is activated or if the
MAX
or
Purge
pushbuttons are
pressed.
Emergency
alternates on/off with the velocity reading.
Fire Alarm
The Fire alarm input is activated.
Fire alarm
alternates on/off with the
velocity reading.
Filter Exp
The Hours Counter is enabled and the set hours limit is exceeded.
Filter
Exp
alternates on/off with the velocity reading. A revolving clock icon
displays when the fan is on and the counter is active.
Содержание AFA4000/E
Страница 1: ...AFA4000 E VAV Airflow Controller Installation and operating manual Issue 2...
Страница 57: ...AFA4000 E VAV Airflow Controller p 57 Figure 22 AFA4000 E inverter control connections...
Страница 58: ...AFA4000 E VAV Airflow Controller p 58 Figure 23 AFA4000 E damper control connections...
Страница 59: ...AFA4000 E VAV Airflow Controller p 59 Figure 24 AFA4000 E typical connection diagram...
Страница 60: ...AFA4000 E VAV Airflow Controller p 60 Figure 25 AFA4000 E AS typical connection diagram...
Страница 62: ...AFA4000 E VAV Airflow Controller p 62 Figure 27 SM7 airflow sensor dimensions...
Страница 71: ...AFA4000 E VAV Airflow Controller p 71 Figure 34 Side views of a rack and pinion drive typical installation...
Страница 96: ...AFA4000 E VAV Airflow Controller p 96 6 9 Auto Sash System Control Box general wiring diagram...
Страница 158: ...AFA4000 E VAV Airflow Controller p 158...
Страница 182: ...AFA4000 E VAV Airflow Controller p 182 Figure 79 Connection diagram with RS232 485 converter...