Function
Description
Emergency
When input configured as Emergency is activated:
•
Red Alarm LED Halo illuminated
•
Emergency
toggles on /off in the status window (when
configured)
•
Audible alarm sounds. Can be muted by pressing the
Mute
pushbutton
•
Emergency alarm relay operates (when configured)
•
Unit is driven to the VAV MAX operating position
Sash High
When input configured as Sash High is activated:
•
Amber Alarm LED Halo illuminated
•
Sash High
toggles on / off in the status window if sash height >
calibrated position
•
Audible alarm sounds. Can be muted by pressing the
Mute
pushbutton
•
Sash High relay operates (when configured)
•
Alarm re-activates after time delay when repeat time enabled
High / Low
When input configured as High/Low is activated:
•
Display Icon shows
↑
(High) or
↓
(Low)
•
High / Low relay operates (when configured)
•
This function is designed for two-speed fan operation or two-
position damper operation switched using a micro switch or
proximity switch activated at a given position on the sash.
Sash Warning
When the input configured as Close Sash is activated:
•
Red Alarm LED Halo illuminated
•
Close Sash
- toggles on/off in the status window
•
Audible alarm sounds (after pre-set time)
•
Audible can be muted by pressing the
Mute
pushbutton. This
silences the alarm (when configured)
•
Audible alarm re-sounds after time delay when repeat time
enabled
•
Reset when the sash is lowered to closed position and the input is
de-activated or an operator is present
Содержание AFA4000/E
Страница 1: ...AFA4000 E VAV Airflow Controller Installation and operating manual Issue 2...
Страница 57: ...AFA4000 E VAV Airflow Controller p 57 Figure 22 AFA4000 E inverter control connections...
Страница 58: ...AFA4000 E VAV Airflow Controller p 58 Figure 23 AFA4000 E damper control connections...
Страница 59: ...AFA4000 E VAV Airflow Controller p 59 Figure 24 AFA4000 E typical connection diagram...
Страница 60: ...AFA4000 E VAV Airflow Controller p 60 Figure 25 AFA4000 E AS typical connection diagram...
Страница 62: ...AFA4000 E VAV Airflow Controller p 62 Figure 27 SM7 airflow sensor dimensions...
Страница 71: ...AFA4000 E VAV Airflow Controller p 71 Figure 34 Side views of a rack and pinion drive typical installation...
Страница 96: ...AFA4000 E VAV Airflow Controller p 96 6 9 Auto Sash System Control Box general wiring diagram...
Страница 158: ...AFA4000 E VAV Airflow Controller p 158...
Страница 182: ...AFA4000 E VAV Airflow Controller p 182 Figure 79 Connection diagram with RS232 485 converter...