AFA4000/E
VAV Airflow Controller / p.110
The lower, middle and top positions are now calibrated.
Press
Enter
to quit calibration mode.
17.
Press
Enter
, to exit Calibration mode.
18.
Press
↓
button, select
Back to Set up
menu.
19.
Press
Enter
.
20.
Press
↓
button, select
Exit and Save Changes.
21.
Press
Enter
.
The Controller will now enter
Run
mode.
Note, the Auto Lock, Auto Stop, Auto Open and Overcurrent Monitor functions also operate in
the DDA profile.
7.6.4
Calibrating the sash lock function
To define the required Sash Lock position:
1.
Press and
hold the
Enter
button for 5 seconds until the
Main Menu
is displayed.
2.
The current Hardware and Software versions are displayed, for example
1. HW2.0 FW1.1
3.
Press
↓
button, select
2. Setup
and press
Enter
.
4.
Enter the Password using the
↑
button and
Enter
button, the factory default is 1 0 0 1. In the
Set-Up menu, the controller will display
Keypad Tones
.
5.
Press
↓
button, select
Sash Settings
and press
Enter
. The controller will display
1. S1 Enabled.
6.
Press
↓
button, select
8. Auto Lock
and press
Enter
.
7.
Set to
Enabled
and press
Enter
.
8.
Use the
↓
button to scroll through the menu and select
Lock Alarm
. Press
Enter
.
9.
Press
↑/↓
buttons to set the required alarm delay time and press
Enter
.
10.
P Use the
↓
button to scroll through the menu and select
Lock Pos Calibration
. Press
Enter
.
11.
Instructions will be displayed, for example: Lock the sash then move the sash to the lowest
lock position. Enter to accept. Cancel to abort
12.
Open the sash to the Lock position and press
Enter
.
If the position chosen is too low, lower than the calibrated Bottom, Middle or Top positions,
Bad position - try again
will be displayed.
If the chosen position is acceptable, the lock position will be set.
13.
Press
↓
button, select
Back to Set up menu
and press
Enter
.
14.
Press
↓
button, select
Exit and Save Changes
and press
Enter
.
The Auto Sash Controller will now go to
RUN
mode.
Содержание AFA4000/E
Страница 1: ...AFA4000 E VAV Airflow Controller Installation and operating manual Issue 2...
Страница 57: ...AFA4000 E VAV Airflow Controller p 57 Figure 22 AFA4000 E inverter control connections...
Страница 58: ...AFA4000 E VAV Airflow Controller p 58 Figure 23 AFA4000 E damper control connections...
Страница 59: ...AFA4000 E VAV Airflow Controller p 59 Figure 24 AFA4000 E typical connection diagram...
Страница 60: ...AFA4000 E VAV Airflow Controller p 60 Figure 25 AFA4000 E AS typical connection diagram...
Страница 62: ...AFA4000 E VAV Airflow Controller p 62 Figure 27 SM7 airflow sensor dimensions...
Страница 71: ...AFA4000 E VAV Airflow Controller p 71 Figure 34 Side views of a rack and pinion drive typical installation...
Страница 96: ...AFA4000 E VAV Airflow Controller p 96 6 9 Auto Sash System Control Box general wiring diagram...
Страница 158: ...AFA4000 E VAV Airflow Controller p 158...
Страница 182: ...AFA4000 E VAV Airflow Controller p 182 Figure 79 Connection diagram with RS232 485 converter...