Performance Verification
4- 18
TLA7AXX Logic Analyzer Module Service Manual
12.
After completing the last probe, you can exit the application or return to the
main window to perform other procedures.
CAUTION.
Power down the test fixture before disconnecting the PIC from the test
fixture. Failure to do so can damage the test fixture or the PIC.
These procedures verify the input resistance, the digital gain and offset, and the
analog gain and offset parameters of the logic analyzer probe input sections.
These procedures are performed without the logic analyzer probes. The
procedures are optional and do not verify any advertised specifications.
However, they are helpful in verifying the logic analyzer functionality and for
troubleshooting purposes.
Equipment Setups.
It is assumed that the logic analyzer is already connected to
the TLACAL2 test fixture and that the performance verification software is
already installed on the hard disk of the logic analyzer. It is also assumed that the
logic analyzer module is installed in the logic analyzer mainframe. The
following procedures describe the specific connections from the logic analyzer
module to the MICs and to the TLACAL2 test fixture.
CAUTION.
If you have any P6041cables connected from the logic analyzer to the
TLACAL2 test fixture, always disconnect one end of each P6041 cable before
powering down the test fixture.
Always power down the test fixture before connecting or disconnecting ribbon
cables. Failure to do so can damage the test fixture, the PIC, or the MICs.
1.
Power down the TLACAL2 test fixture before connecting the ribbon cables
and the MICs.
CAUTION.
To avoid damaging the test fixture, the PIC, or the MICs through
static discharge, always wear a grounded antistatic wrist strap. Avoid handling
individual components of the test fixture without static protection.
2.
Connect the four ribbon cables to the ribbon cable connectors on the test
fixture.
3.
Disconnect the PIC if it is connected to the ribbon cable.
4.
Connect a MIC to each of the ribbon cables.
5.
Connect any one of the coaxial cables from the CS (current source)
connectors on the test fixture to the CS connector on the MIC.
MIC Procedures
Содержание TLA7AA1
Страница 5: ......
Страница 17: ...Service Safety Summary xii TLA7AXX Logic Analyzer Module Service Manual...
Страница 23: ...Preface xviii TLA7AXX Logic Analyzer Module Service Manual...
Страница 27: ...Introduction xxii TLA7AXX Logic Analyzer Module Service Manual...
Страница 63: ...Operating Information 2 20 TLA7AXX Logic Analyzer Module Service Manual...
Страница 105: ...Adjustment Procedures 5 4 TLA7AXX Logic Analyzer Module Service Manual Figure 5 1 TLACAL2 test fixture connections...
Страница 157: ...Troubleshooting 6 38 TLA7AXX Logic Analyzer Module Service Manual...
Страница 163: ...Electrical Parts List 8 2 TLA7AXX Logic Analyzer Module Service Manual...