Adjustment Procedures
TLA7AXX Logic Analyzer Module Service Manual
5- 7
b.
Connect the clock/qualifier leads to the SNGL square pins at J220. Make
sure that you connect the ground side or the “--” side to the square pins
with the GND label. There is no need to connect the single ground wire.
8.
If you are adjusting a P6860, connect the probe heads to the P6860/80 GAIN
& OFFSET connections at J120 and J130 on the test fixture. If you have a
P6880 probe, connect the probe heads to J120, J121, J130, and J131 on the
test fixture.
9.
Power on the test fixture, logic analyzer, and digital multimeter and allow all
test equipment to warm up for 30 minutes before continuing the adjustments.
Complete the following steps to start the performance verification software and
to run the adjustments:
1.
Exit the TLA application on the logic analyzer.
2.
Double-click the TLA7Axx Performance Verification icon on the desktop.
An application window similar to Figure 5--2 appears.
Figure 5- 2: Performance verification software startup window
3.
Click the Adjustment button to start the software. An Adjustment Procedure
dialog box similar to Figure 5--3 appears.
Start the Software
Содержание TLA7AA1
Страница 5: ......
Страница 17: ...Service Safety Summary xii TLA7AXX Logic Analyzer Module Service Manual...
Страница 23: ...Preface xviii TLA7AXX Logic Analyzer Module Service Manual...
Страница 27: ...Introduction xxii TLA7AXX Logic Analyzer Module Service Manual...
Страница 63: ...Operating Information 2 20 TLA7AXX Logic Analyzer Module Service Manual...
Страница 105: ...Adjustment Procedures 5 4 TLA7AXX Logic Analyzer Module Service Manual Figure 5 1 TLACAL2 test fixture connections...
Страница 157: ...Troubleshooting 6 38 TLA7AXX Logic Analyzer Module Service Manual...
Страница 163: ...Electrical Parts List 8 2 TLA7AXX Logic Analyzer Module Service Manual...