Adjustment Procedures
5- 16
TLA7AXX Logic Analyzer Module Service Manual
17.
Click the Finish button to finish the procedure and return to the startup
window.
Merged Deskew Procedures
The Merged Deskew procedures adjusts the deskew between modules in a
merged module set. These procedures use the logic analyzer probes connected to
the Deskew and Minimum Pulse connections on the TLACAL2 test fixture.
Depending on the type of probe that you are using and on the number of
channels of your logic analyzer module, the test connections differ. You should
run these procedures after you have completed the individual module merge
procedure.
These procedures require similar setups as described in the Performance
Verification chapter of this manual. It is assumed that the logic analyzer is
already connected to the test fixture and that the performance verification
software is already installed on the hard disk of the logic analyzer. It is also
assumed that the logic analyzer modules are installed in the logic analyzer
mainframe. The following procedures describe the specific connections from the
logic analyzer modules to the probes and the TLACAL2 test fixture.
These procedures adjust one pair of modules at a time. When connecting the
probes to the modules, make sure that you know which module is the master
module and which module is the slave module.
1.
Connect a probe from the C0/C1/C2/C3 input connector (or the
C3/C2/A3/A2 connector) on the slave module to the appropriate DATA
connector as listed in Table 5--3.
2.
Connect a probe from the C0/C1/C2/C3 input connector (or the
C3/C2/A3/A2 connector) on the master module to the appropriate DESKEW
REF connector as listed in Table 5--3.
NOTE
. If you are using P6860 probes or P6880 probes, make sure that the
alignment pin on the probe head is aligned with the hole on the test fixture.
Tighten the probe head screws by alternating between them until they are finger
tight (no more than 1 in-lbs of torque).
3.
If you are using a P6810 probe, make sure that you connect the ground or
low side of the lead sets to the GND side of the connectors on the test
fixture. Connect the clock/qualifier leads to the DESKEW REF (J880)
connector; make sure that you connect the ground side or the “--” side to the
square pins with the GND label. There is no need to connect the single
ground wire.
Equipment Setups
Содержание TLA7AA1
Страница 5: ......
Страница 17: ...Service Safety Summary xii TLA7AXX Logic Analyzer Module Service Manual...
Страница 23: ...Preface xviii TLA7AXX Logic Analyzer Module Service Manual...
Страница 27: ...Introduction xxii TLA7AXX Logic Analyzer Module Service Manual...
Страница 63: ...Operating Information 2 20 TLA7AXX Logic Analyzer Module Service Manual...
Страница 105: ...Adjustment Procedures 5 4 TLA7AXX Logic Analyzer Module Service Manual Figure 5 1 TLACAL2 test fixture connections...
Страница 157: ...Troubleshooting 6 38 TLA7AXX Logic Analyzer Module Service Manual...
Страница 163: ...Electrical Parts List 8 2 TLA7AXX Logic Analyzer Module Service Manual...