Performance Verification
TLA7AXX Logic Analyzer Module Service Manual
4- 3
Test Equipment
The procedures use external, traceable signal sources to directly test characteris-
tics that are designated as checked (
n
) in the
Specifications
chapter of this
manual. Table 4--1 shows the required equipment list; this equipment is required
for the performance verification and adjustment procedures.
Table 4- 1: Test equipment
Item number and description
Minimum requirements
Example
1. Mainframe
TLA700 Series Mainframe with TLA Applica-
tion software V4.2 or higher
TLA721 Benchtop Mainframe or
TLA715 Portable Mainframe
2. Logic analyzer module with
probes
TLA7Axx Series logic analyzer module
No substitute allowed
3. Logic analyzer probes
P6810, P6860, or P6880 logic analyzer
probes. One probe required for every
34-channels on the logic analyzer module
No substitute allowed
4. Performance Verification test
fixture with accessories
TLACAL2 test fixture; Includes:
4 Ribbon cables
4 Module interface connectors (MICs)
1 Probe interface connector (PIC)
4 P6041 BNC-to-SMB cables
8 MCX-to-MCX coaxial cables, with
identification color bands
No substitute allowed
5. Digital Multimeter with probes
1
Agilent 34401A, 6.5 digit display, 35 ppm, 1
year accuracy, 1000 readings per minute,
100 nV sensitivity, GPIB controllable
No substitute allowed
6. Oscilloscope
2
Tektronix TDS694C Four channels, 3 GHz,
10GS/s
No substitute allowed
7. USB-GPIB controller:
National Instruments 778195-01
Windows 2000 compatible with 2 m cable.
Tektronix iView cable
8. GPIB cable
2 m GPIB cable.
Tektronix part number 012-0991-00
9. RS-232 cable
2 m RS-232 cable, 9-pin female-to-female
connector
Tektronix part number 012-1379-00
10. BNC cables
4 Analog Output BNC-to-BNC, low loss, 33 in.
10x cables
Tektronix part number 174-4595-00
1
The digital multimeter is required to complete the Probe+PIC performance verification and adjustment procedure, the
MIC performance verification and adjustment procedures, the Probe Gain & Offset performance
verification procedures, and the certification procedures.
2
The oscilloscope is required to complete the performance verification procedures for the Analog Output tests.
Содержание TLA7AA1
Страница 5: ......
Страница 17: ...Service Safety Summary xii TLA7AXX Logic Analyzer Module Service Manual...
Страница 23: ...Preface xviii TLA7AXX Logic Analyzer Module Service Manual...
Страница 27: ...Introduction xxii TLA7AXX Logic Analyzer Module Service Manual...
Страница 63: ...Operating Information 2 20 TLA7AXX Logic Analyzer Module Service Manual...