Performance Verification
4- 4
TLA7AXX Logic Analyzer Module Service Manual
Test Equipment Setup
The TLACAL2 Performance Verification and Adjustment fixture and other test
equipment requires preliminary installation and setups. You will need the
equipment listed in Table 4--1 on page 4--3 to complete the performance
verification procedures. If you have not already installed the test fixture and test
equipment, refer to
TLACAL2 Fixture Installation
on page 2--3 for the prelimi-
nary installation instructions.
Most of the individual procedures require specific test equipment setups
depending on your logic analyzer configuration. These setups are listed prior to
each test and are also described in the TLACAL2 software windows.
CAUTION.
Always power down the TLACAL2 test fixture before connecting or
disconnecting the ribbon cables. Failure to do so can damage the test fixture, the
PIC, or the MICs.
The performance verification software is a separate application that consists of
executable software files. The software must be installed on the hard disk before
you can use it; refer to
Software Installation
on page 2--7 for instructions on
installing the software. You must quit the logic analyzer application before
starting the performance verification software; you cannot run both applications
at the same time.
Use the following steps to start and run the software. These steps are repeated
under the individual procedures but are listed here to give you an overview of
using the software.
1.
Allow the logic analyzer and all test equipment to warm up for at least 30
minutes.
2.
Exit the logic analyzer application.
3.
Select Start
→
Programs
→
Tektronix Logic Analyzer
→
TLA7AxxPVAd-
just (or double-click on the TLA7AxxPVAdjust icon on the desktop). An
application window similar to Figure 4--2 appears.
4.
Select the module that you want to test from the list and then click either the
Verification button, Adjustment button, or Certification button depending on
the type of procedure that you want to perform.
Starting the Performance
Verification Software
Содержание TLA7AA1
Страница 5: ......
Страница 17: ...Service Safety Summary xii TLA7AXX Logic Analyzer Module Service Manual...
Страница 23: ...Preface xviii TLA7AXX Logic Analyzer Module Service Manual...
Страница 27: ...Introduction xxii TLA7AXX Logic Analyzer Module Service Manual...
Страница 63: ...Operating Information 2 20 TLA7AXX Logic Analyzer Module Service Manual...
Страница 105: ...Adjustment Procedures 5 4 TLA7AXX Logic Analyzer Module Service Manual Figure 5 1 TLACAL2 test fixture connections...
Страница 157: ...Troubleshooting 6 38 TLA7AXX Logic Analyzer Module Service Manual...
Страница 163: ...Electrical Parts List 8 2 TLA7AXX Logic Analyzer Module Service Manual...