Program
Local/Remote
Mode
99
1 ASCII
0 or 1
6.5 RESPONSE
OVERVIEW
The command responses will follow the same format as outlined above in
section 6.1. This list is comprised of Commands with complex responses
only. Commands using a simple response will use the <$> character
(ASCII 0x24) as a “Success” response or a single character error code.
These responses will be eight ASCII characters in length.
Response Name
<CMD>
Response
Request kV
Setpoint
14
8-11 ASCII
Request mA
Setpoint
15
8-11 ASCII
Request Analog
Monitor
Readbacks
19
12-23
ASCII
Request Total
Hours High
Voltage On
21
14 ASCII
Request Status
22
13 ASCII
Request DSP
Software Version
23
18 ASCII
Request Hardware
Version
24
10 ASCII
Request Web
Server Version
25
18 ASCII
Request Model
number
26
12 ASCII
Request User
Configs
27
9-17 ASCII
Request unit
Scaling
28
24-32
ASCII
Read Interlock
Status
55
8 ASCII
Request kV
monitor
60
8-11
ASCII
Request mA
monitor
61
8-11
ASCII
Request –15V
LVPS
65
8-11
ASCII
118080-001 REV A
Page 70 of 95
Содержание SLM SERIES
Страница 15: ...SLM MANUAL 6 118073 001 Rev C Figure 2 2 Unit Dimensions 1200W ...
Страница 19: ...SLM MANUAL 10 118073 001 Rev C Figure 3 3 Local Programming via External Voltage Source ...
Страница 20: ...SLM MANUAL 11 118073 001 Rev C Figure 3 4 Remote Monitoring ...
Страница 40: ...Figure 9 Web Page 1 Contact Information 118080 001 REV A Page 13 of 95 ...
Страница 46: ...4 2 2 9 About Displays version information and model number Figure 15 About 118080 001 REV A Page 19 of 95 ...
Страница 98: ...Request Faults 68 20 ASCII 118080 001 REV A Page 71 of 95 ...