6.6.24 Request Analog Monitor Readbacks
Description:
The host requests that the firmware transmit the present values of Analog
Monitor Readbacks.
Direction:
Host to supply
Syntax:
<STX><19><,><
CSUM><ETX>
Example:
<STX><19>,<
CSUM><ETX>
Response:
<STX><19><,><ARG1><,><ARG2><,><ARG3><,>< CSUM><ETX>
Where:
ARG1 = kV monitor = 0 – 4095
ARG2 = mA monitor = 0 – 4095
ARG3 = unused = 0– 4095
Example:
<STX><19>,4095,4095,4095,< CSUM><ETX>