![Spellman SLM SERIES Скачать руководство пользователя страница 84](http://html.mh-extra.com/html/spellman/slm-series/slm-series_instruction-manual_1346474084.webp)
5.5.17 Request Faults
Description:
The host requests that the firmware report Faults.
Direction:
Host to supply
Syntax:
<STX><68><,><ETX>
Response:
<STX><68><,><ARG1><,><ARG2><,><ARG3><,><ARG4><,><ARG5><,
><ARG6><,><ARG7><,><ETX>
Where:
<ARGx> 1 = Fault, 0 = No Fault in ASCII format
ARG1 = ARC
ARG2 = Over Temperature
ARG3 = Over Voltage
ARG4 = Under Voltage
ARG5 = Over Current
ARG6 = Under Current
ARG7 = Watchdog
Example:
<STX>68,0,0,0,0,1,0,0,<ETX>