Chapter 6: Alarms
6-4
Veterinary 3 Parameter Advisor
®
Operation Manual
To change the alarm volume:
Turn the rotary knob on the monitor to move the cursor. On the main menu at the bottom of the display,
1.
highlight ALARMS and push the knob to select.
Highlight ALARM VOLUME and push the knob to select.
2.
Turn the rotary knob to increase or decrease the volume to the desired level and push the knob to select. A
3.
momentary beep tone will sound at each new level of volume.
Highlight MAIN MENU and push the knob to select.
4.
Stop alarms automatically or manually
Your monitor will either be set to automatically stop an alarm as soon as the measured value returns to within
the alarm limits, or it will be set so you must acknowledge the alarm manually. See
Select the Alarm Reset
Operation
in
Chapter 12: Service Menu
for instructions for choosing AUTO RESET or MANUAL RESET.
If your monitor is set to AUTO RESET, all alarm indicators for a parameter will cease as soon as the measured value
is once again within the alarm limits.
If your monitor is set to MANUAL RESET, you must acknowledge an alarm condition by pressing the alarm silence
key (
B
). If the measured value returns to within the alarm limits before you press the alarm silence key, a
flashing up or down arrow will be displayed next to the measured value to indicate the violation and the
auditory alarm will continue until you press the alarm silence key.
Silence alarms
If your monitor is set to allow TEMPORARY OR INDEFINITE audio alarm silence, then pressing the ALARM SILENCE
key while audible alarms are enabled will do one of the following:
Press ALARM SILENCE momentarily to silence alarms for two minutes. AUDIO PAUSED will appear at the
•
top of the display with a two-minute timer. Audible alarms will be re-enabled if any new alarm occurs, if
the two-minute timer times out, or if the ALARM SILENCE key is pressed again.
Press and hold the Alarm Silence key for about three (3) seconds to silence alarms indefinitely. AUDIO
•
PAUSED will be displayed with no timer. Audible alarms will be re-enabled if any new alarm occurs, or if
the ALARM SILENCE key is pressed again.
Press and hold ALARM SILENCE for six (6) or more seconds to silence alarms permanently. AUDIO OFF will
•
be displayed. Audible alarms will NOT be re-enabled if any new alarm occurs. Press the ALARM SILENCE
key again to re-enable audible alarms.
If audio alarm silence for your monitor is set to allow TEMPORARY ONLY, then pressing the ALARM SILENCE key
will silence alarms for two minutes. AUDIO PAUSED will be displayed with a two-minute timer (2:00). Audible
alarms will be re-enabled if any new alarm occurs, if the two-minute timer times out, or if the ALARM SILENCE
key is pressed again.
If audio alarm silence for your monitor is set to allow INDEFINITE ONLY, then pressing and holding ALARM
SILENCE key will silence alarms for an unspecified amount of time. AUDIO PAUSED will be displayed with no
timer. The audible alarms will be re-enabled if any new alarm occurs, or it the ALARM SILENCE key is pressed
again.
NOTE! Only qualified personnel may change the allowed alarm audio silence mode. See
Select Alarm
Audio Silence Operation
in
Chapter 12: Service Menu
for details regarding alarm silence.
Refer to the chapters dedicated to each parameter for details regarding the high, medium, and low alarming
conditions.
Содержание SurgiVet Advisor WWV9230
Страница 2: ......
Страница 10: ...Table of Contents viii Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 68: ...Chapter 7 ECG 7 14 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 80: ...Chapter 8 Oximetry 8 12 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 106: ...Chapter 12 Service Menu 12 8 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 120: ...Appendix D Revision History D 2 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 121: ......