Chapter 9: Non-invasive Blood Pressure
Veterinary 3 Parameter Advisor
®
Operation Manual
9-7
Choose the Inflation Pressure
You can change the cuff inflation pressure before any measurement. If you change the pressure, the monitor
will use the new value for the next NIBP measurement. The default settings are 200 mmHg for the cat mode, 150
mmHg for the dog mode, and 100 mmHg for the horse mode.
After an NIBP measurement is taken and there is a valid systolic pressure reading displayed in the NIBP
parameter box, the cuff inflation pressure will automatically inflate to the systolic pressure plus 35 mmHg for
cat and dog patient types, and plus 20 for the horse type. For example, if the latest NIBP measurement for a cat
shows a systolic pressure of 112 mmHg, then the next time a measurement is taken the cuff inflation pressure
will be 147 mmHg (unless you change it).
To choose the cuff inflation pressure:
On the NIBP parameter menu, highlight INFL PRESS and push the knob to select.
1.
Turn the rotary knob to choose the desired pressure (50-250 mmHg for cat and dog patients, 50-120 for
2.
horses) and push the knob to select.
Highlight MAIN MENU and push the knob to select.
3.
Adjust the Volume of the NIBP Completion Beep Tone
The monitor will issue a beep tone when each NIBP measurement has been completed. You can control the
volume of the beep tone.
To adjust the volume of the NIBP completion beep:
On the NIBP parameter menu, highlight COMPLETION VOLUME and push the knob to select.
1.
Turn the rotary knob to increase or decrease the volume to the desired level and push the knob to select. A
2.
momentary beep tone will sound at each new level of volume.
Highlight MAIN MENU and push the knob to select.
3.
Measuring Non-invasive Blood Pressure (NIBP)
Manual NIBP Mode
If you selected MANUAL as the NIBP MODE, press the NIBP key (
h
) on the monitor.
If the measurement is successful, you will see the systolic, diastolic, and mean arterial pressure values displayed
in the NIBP parameter box and you will hear the completion beep. The time of the successful reading will also be
displayed in the parameter box.
If the measurement is not successful, the previous pressure values will be displayed. If there is no previous data,
you will see dashes (- - -) displayed in the place of the pressure values. The monitor will attempt to obtain a
reading for up to 120 seconds for cat and dog patient types. For the horse patient type, the monitor will retry
for 90 seconds. If the monitor is unable to take a measurement, you will hear a bell tone and see a message
displayed in the parameter box. To acknowledge the NIBP error message, press the NIBP key (
h
) on the
monitor while the message is displayed in the NIBP parameter box.
Содержание SurgiVet Advisor WWV9230
Страница 2: ......
Страница 10: ...Table of Contents viii Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 68: ...Chapter 7 ECG 7 14 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 80: ...Chapter 8 Oximetry 8 12 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 106: ...Chapter 12 Service Menu 12 8 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 120: ...Appendix D Revision History D 2 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 121: ......