Chapter 12: Service Menu
12-4
Veterinary 3 Parameter Advisor
®
Operation Manual
Select the Alarm Audio Silence Operation
You can choose how the alarm silence key (
B
) will function.
If you select TEMPORARY ONLY:
When you press the alarm silence key (
•
B
), an audible alarm will be temporarily silenced for two minutes.
AUDIO PAUSED will be displayed on the upper right side of the display. The audible alarm will be re-
enabled if any new alarm condition occurs, or in two minutes if no new alarm condition occurs.
If you select INDEFINITE ONLY:
When you press the alarm silence key (
•
B
) for three seconds, audible alarms will be indefinitely silenced.
AUDIO PAUSED will be displayed on the upper right side of the display. If any new alarm condition occurs
while audible alarms are indefinitely silenced, the audible alarm will be re-enabled.
If you select TEMPORARY AND INDEFINITE:
When you press and hold the alarm silence key (
•
B
) for six seconds, audible alarms will be permanently
silenced. AUDIO OFF will be displayed on the upper right side of the display. The audible alarm will not be
re-enabled if a new alarm condition occurs.
NOTE! This option is only available if Dual Alarm Silence Allowed is enabled in the factory service menu.
To select the alarm audio silence operation:
On the service menu, highlight MONITOR DEFAULTS and push the knob to access the submenu for default
1.
parameter alarm limits.
Highlight ALARM AUDIO SILENCE OPERATION and push the knob to select.
2.
Turn the rotary knob on the monitor to choose the desired setting (TEMPORARY ONLY, INDEFINITE ONLY, or
3.
TEMPORARY AND INDEFINITE) and push the knob to select.
Highlight MAIN or PREVIOUS and push the knob to select.
4.
Select the Alarm Reset Operation
You can set the monitor to automatically silence an audible alarm tone as soon as the measured value returns to
within the alarm limits (AUTO), or you can set it so that an audible alarm can be manually silenced (MANUAL) by
pressing the alarm silence key (
B
).
To select the alarm reset operation:
On the service menu, highlight MONITOR DEFAULTS and push the knob to access the submenu for default
1.
parameter alarm limits.
Highlight RESET ALARMS and push the knob to select.
2.
Turn the rotary knob on the monitor to choose the desired setting (AUTO or MANUAL) and push the knob to
3.
select.
Highlight MAIN or PREVIOUS and push the knob to select.
4.
Содержание SurgiVet Advisor WWV9230
Страница 2: ......
Страница 10: ...Table of Contents viii Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 68: ...Chapter 7 ECG 7 14 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 80: ...Chapter 8 Oximetry 8 12 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 106: ...Chapter 12 Service Menu 12 8 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 120: ...Appendix D Revision History D 2 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 121: ......