Chapter 8: Oximetry
8-4
Veterinary 3 Parameter Advisor
®
Operation Manual
Attaching the Patient
WARNING! Prolonged use or the patient’s condition may require changing the sensor site periodically.
Change the sensor site and check skin integrity, circulatory status, and correct alignment at
least every 4 hours.
WARNING! When attaching sensors with Microfoam
®
tape, do not stretch the tape or attach the tape too
tightly. Tape applied too tightly may cause inaccurate readings and blisters on the patient’s
skin (lack of skin respiration, not heat, causes the blisters).
Choose a sensor to be used to monitor oximetry.
1.
SURGIVET
®
SENSOR SELECTION CHART
Patient
Site
Description
Small/Medium Animal
Ear, Toe webbing, Tongue, Thin Tissue
Rectum/Tail
Hock, Achilles tendon, Thicker Tissue
V3078: Mini Clip
V1700: Reflectance Sensor
V1707: Universal ‘C’ Sensor
Large Animal
Rectum/Tail
Hock, Achilles tendon, Thicker Tissue
V1700: Reflectance Sensor
V1707 Universal ‘C’ Sensor
Horse
Tongue
V1707 Universal ‘C’ Sensor
Clean and disinfect the sensor. Use a soft cloth moistened in water or a mild soap solution, and then wipe
2.
the sensor with isopropyl alcohol.
WARNING! Do not autoclave, ethylene oxide sterilize, or immerse the sensors in liquid. Evidence that
liquid has been allowed to enter the monitor voids the warranty.
CAUTION! Unplug the sensor from the monitor before cleaning or disinfecting.
NOTE! Obstructions or dirt on the sensor’s red light or detector may cause a sensor failure. Make sure
there are not obstructions and the sensor is clean.
Check the sensor and the patient cable to make sure they do not appear damaged.
3.
WARNING! Using a damaged sensor may cause inaccurate readings, possibly resulting in patient injury
or death. Inspect each sensor. If a sensor appears damaged, do not use it. Use another sensor
or contact your authorized repair center for help.
WARNING! Using a damaged patient cable may cause inaccurate readings, possibly resulting in patient
injury or death. Inspect the patient cable. If the patient cable appears damaged, do not use it.
Contact your authorized repair center for help.
WARNING! If any of the integrity checks fail, do not attempt to monitor the patient. Use another sensor
or patient cable, or contact the equipment dealer for help if necessary.
Содержание SurgiVet Advisor WWV9230
Страница 2: ......
Страница 10: ...Table of Contents viii Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 68: ...Chapter 7 ECG 7 14 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 80: ...Chapter 8 Oximetry 8 12 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 106: ...Chapter 12 Service Menu 12 8 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 120: ...Appendix D Revision History D 2 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 121: ......