Chapter 12: Service Menu
12-6
Veterinary 3 Parameter Advisor
®
Operation Manual
When the static limits feature is on and you change high and low alarm limits in the alarm limits box, or the
default alarm limits box, or you restore the factory default alarm limits, these limits will be applied immediately. If
you change the patient type or turn off the monitor and then turn it on again, the changes in the alarm limits will
be retained in the alarm limit box and the default alarm limit box.
Note that when the static limits feature is on or off, only the parameters that are enabled, or turned on, will be
displayed in the alarm limits box. All of the parameters and their alarm limits are displayed in the default alarm
limit box.
The default setting for the static limits feature is ON.
To turn the static limits feature on or off:
On the service menu, highlight STATIC LIMITS and push the knob to select.
1.
Highlight YES or NO to cancel and push the knob to select.
2.
Highlight MAIN or PREVIOUS and push the knob to select.
3.
Using the Factory Service Menu
This menu is intended for service personnel. Contact your authorized repair center for help.
Using the Demo Mode
WARNING! While the demo mode is active, no patient data is collected or analyzed. Never attach a
patient to the monitor while it is in demo mode.
The Advisor
®
includes a demonstration mode to be used for training and sales activities. Installed parameters,
including non-invasive blood pressure, are simulated when the demo mode is turned on. All of the functions of
the monitor will be simulated in the demo mode, including alarms, trends, auto print, and NIBP history.
To turn on the demo mode:
On the service menu, highlight DEMO MODE and push the knob to select.
1.
Turn the rotary knob to highlight YES and push the knob to select.
2.
DEMO will be displayed, red and flashing, in the message area in the lower right corner of the display.
•
To turn off the demo mode:
On the service menu, highlight DEMO MODE and push the knob to select.
1.
Turn the rotary knob to highlight NO and push the knob to select.
2.
The monitor will return to normal operating conditions and will collect patient data. When the demo
•
mode is turned off, the demo options menu is not available from the setup menu.
Содержание SurgiVet Advisor WWV9230
Страница 2: ......
Страница 10: ...Table of Contents viii Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 68: ...Chapter 7 ECG 7 14 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 80: ...Chapter 8 Oximetry 8 12 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 106: ...Chapter 12 Service Menu 12 8 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 120: ...Appendix D Revision History D 2 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 121: ......