Chapter 4: Setting up the Monitor
Veterinary 3 Parameter Advisor
®
Operation Manual
4-1
Unpack the Monitor and Check the Shipment
Before you begin setting up, carefully remove the Advisor
®
Vital Signs Monitor and its accessories from the
shipping carton. Save the packing materials in case the monitor must be shipped or stored. Compare the
packing list with the supplies and equipment you received to make sure the shipment is complete.
Quick Setup Instructions
Follow these setup steps every time you begin monitoring a patient. See
Detailed Setup Instructions
in this
chapter for a thorough explanation of each step.
Choose the installation site.
1.
Check the monitor’s AC rating.
2.
Connect the AC power cord.
3.
Press the ON/OFF key
4.
x
to turn on the monitor.
Select the patient type.
5.
Add a new patient.
6.
Use the SETUP menu to define some general configuration information and parameter options.
7.
If necessary, set the time and date on the display.
8.
Detailed Setup Instructions
Set up the monitor in a room with a temperature of 0-40° C (32-104° F) and a relative humidity of 15-95%,
1.
non-condensing.
If the monitor was in an area having a temperature higher or lower than this, wait a few minutes before
a.
setting up and using the monitor.
Check the monitor’s AC rating. Check the monitor’s AC rating plate to ensure the nominal voltage at your
2.
installation site matches the monitor’s rating.
If the AC rating is not correct, do not use the monitor. Contact your authorized repair center for help.
a.
Plug the AC power cord into the power connector on the back of the monitor.
3.
Plug the other end of the AC power cord into a grounded, three-wire hospital-grade outlet.
4.
Verify that the front panel AC Power LED
5.
X
is lit.
WARNING! Do not plug the monitor into an outlet controlled by a wall switch.
Press the ON/OFF key
6.
x
to turn on the monitor.
The display will light up, the monitor will begin a brief system check, and then automatically enter the
a.
monitoring mode.
NOTE! The monitor performs a number of system checks during its start-up time. If the monitor detects
an error with its internal circuitry, a message is displayed and the monitor will not enter the
monitoring mode.
Chapter 4: Setting up the Monitor
Содержание SurgiVet Advisor WWV9230
Страница 2: ......
Страница 10: ...Table of Contents viii Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 68: ...Chapter 7 ECG 7 14 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 80: ...Chapter 8 Oximetry 8 12 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 106: ...Chapter 12 Service Menu 12 8 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 120: ...Appendix D Revision History D 2 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 121: ......