Chapter 5: Monitoring the Patient
Veterinary 3 Parameter Advisor
®
Operation Manual
5-3
Turn the rotary knob to move the cursor back to the name of the measured value and push to select.
9.
Highlight MENU in the alarm limit box and push the knob to select.
10.
Highlight MAIN MENU and push the knob to select.
11.
Use these Features as Needed
Responding to an Alarm
When a numeric measured value matches or exceeds the high or low limit set for that parameter, an alarm
1.
is issued. An audible alarm tone will sound, ALARM will appear in the alarm status line, and the violating
measured value and violated alarm limit will flash in the parameter box.
The alarming action will cease when the measured value is once again within the alarm limits. Your monitor
2.
will either be set to automatically stop alarming as soon as the measured value returns to within the alarm
limits, or it will require you to manually acknowledge the alarm by pressing the alarm silence key (
B
). See
Select the Alarm Reset Operation
in
Chapter 12: Service Menu
for instructions for choosing AUTO RESET or
MANUAL RESET.
NOTE! Only qualified personnel may set the automatic or manual alarm silence, or enable indefinite
alarm silence. See
Chapter 12: Service Menu
for details regarding alarm reset and silence.
For MANUAL RESET alarms, if the measured value returns to within the alarm limits before you press the
3.
alarm silence key, a flashing up (
) or down (
) arrow will be displayed next to the measured value to
indicate the violation and the auditory alarm will continue until you press the alarm silence key.
If the alarm limit is still violated after two minutes, the audible alarm tone will sound again. If, within the two
4.
minutes of alarm silence, another measured value matches or exceeds its alarm limits, the alarming action
will resume, including the audible alarm tone.
If it is appropriate for your patient, you can turn off the alarm detection capability for a single parameter so
5.
that when the measured value matches or exceeds the alarm limits, the monitor will not issue an alarm.
Содержание SurgiVet Advisor WWV9230
Страница 2: ......
Страница 10: ...Table of Contents viii Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 68: ...Chapter 7 ECG 7 14 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 80: ...Chapter 8 Oximetry 8 12 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 106: ...Chapter 12 Service Menu 12 8 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 120: ...Appendix D Revision History D 2 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 121: ......