Chapter 5: Monitoring the Patient
Veterinary 3 Parameter Advisor
®
Operation Manual
5-1
Follow the steps in
Chapter 4: Setting Up the Monitor
; the remainder of this chapter assumes that the monitor is
properly installed and set up.
General Monitoring Instructions
Regardless of the parameters or measured values you want to monitor, follow these steps when you are ready to
attach a patient. Each step is further explained in this chapter.
Attach the patient and sensors.
1.
Choose the waveforms to be displayed.
2.
Adjust the settings in the parameter boxes.
3.
Set the high and low alarm limits.
4.
Use these features as needed:
5.
Silence an alarm
•
Use QUICKSET alarms
•
NIBP mode
•
Suspend mode
•
Freeze mode
•
Trends
•
Attach the Patient
Attach the patient to the desired sensors and connect the sensor cables to the monitor.
Some parameters will automatically turn on when the sensor cable is connected to the monitor and the related
parameter box will appear on the display. Others require you to turn on the parameter box using the SETUP
menu and the PARAMETER OPTIONS submenu. Refer to the chapters dedicated to each parameter for details
regarding sensors and specific parameter box functions.
Adjust the Waveforms Settings
Choose the waveforms to be displayed in positions 2-4 and adjust the settings for each waveform.
Turn the rotary knob on the monitor to move the cursor. Highlight the waveform label and push the knob
to access the waveform menu in the lower left corner of the display. The settings available for the selected
waveform will be displayed.
Waveform 1 is dedicated to the primary ECG lead, although you can specify which lead is to be the
•
primary.
Waveform 2 defaults to the SpO
•
2
waveform. It can be assigned to another ECG lead (in five-lead mode
only), a cascade (continuation) of the primary ECG lead waveform, or a waveform from any other enabled
parameter.
Waveform 3 can be changed to display a cascade (continuation) of the primary ECG lead waveform after
•
waveforms 1 and 2, or it can be a waveform from any other enabled parameter.
Waveform 4 can be changed to display a cascade (continuation) of the primary ECG lead waveform after
•
waveforms 1, 2, and 3, or it can be a waveform from any other enabled parameter.
Refer to the chapters dedicated to each parameter for more details regarding waveform settings.
Chapter 5: Monitoring the Patient
Содержание SurgiVet Advisor WWV9230
Страница 2: ......
Страница 10: ...Table of Contents viii Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 68: ...Chapter 7 ECG 7 14 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 80: ...Chapter 8 Oximetry 8 12 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 106: ...Chapter 12 Service Menu 12 8 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 120: ...Appendix D Revision History D 2 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 121: ......