Chapter 8: Oximetry
Veterinary 3 Parameter Advisor
®
Operation Manual
8-11
Oximetry Alarm Messages
The following messages can be displayed in the SpO
2
parameter box:
ALARM MESSAGE
POSSIBLE CAUSE
CORRECTIVE ACTION
CHECK SENSOR
This message is displayed in the SpO
2
parameter box when the monitor is
unable to detect that the sensor is
connected to the patient. This is a
medium priority alarm.
Be sure the sensor is properly
positioned on the patient.
Be sure that the patient cable is
connected to the monitor.
PULSE SEARCH
This message is displayed in the SpO
2
parameter box when the oximeter
circuitry is optimizing the LED drive
levels in the sensor to properly detect
the PPR and %SpO
2
measured values.
Be sure the sensor is properly
positioned on the patient.
SEARCHING TOO LONG
This message is displayed in the SpO
2
parameter box when the monitor
detects that an SpO
2
sensor is attached
to the patient, but cannot find a pulse
after searching 15-20 seconds. This is a
medium priority alarm.
Check the patient’s condition, and
then be sure the sensor is properly
positioned.
SMALL SIGNAL
This message is displayed in the SpO
2
parameter box when the pulse bar
graph in the parameter box does not
move higher than two bar segments.
Reposition the sensor on the patient.
OXIMETER ERROR
This message is displayed in the SpO
2
parameter box when the oximetry
processor detects an error in the
oximeter’s circuitry. All oximeter
functions are disabled. This is a medium
priority alarm.
Turn the monitor off and back on
again.
Contact your authorized repair center
for help.
COM ERROR
This message is displayed in the SpO
2
parameter box when the monitor’s
main processor cannot communicate
with the oximeter. All oximeter
functions are disabled. This is a medium
priority alarm.
Turn the monitor off and back on
again.
Contact your authorized repair center
for help.
Содержание SurgiVet Advisor WWV9230
Страница 2: ......
Страница 10: ...Table of Contents viii Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 68: ...Chapter 7 ECG 7 14 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 80: ...Chapter 8 Oximetry 8 12 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 106: ...Chapter 12 Service Menu 12 8 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 120: ...Appendix D Revision History D 2 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 121: ......