Chapter 13: Specifications
Veterinary 3 Parameter Advisor
®
Operation Manual
13-3
NIBP
Blood Pressure Measuring
Method of Measurement: Oscillometric with step-down deflation
Range: Systolic: 40 to 265 mmHg
Mean Arterial: 27 to 222 mmHg
Diastolic: 20 to 200 mmHg
Pulse: 25 to 300 bpm
NIBP Accuracy:
Algorithm based on human algorithm that meets or exceeds accuracy
requirements of ANSI / AAMI SP10:1992 and 2002 standards for
non-invasive blood pressure using the oscillometric method.
(AAMI Standard: ± 5 mmHg mean error, 8 mmHg standard deviation)
Inflation Pressure Settings
Adjustment Range: 60 to 280 mmHg – Cat
120 to 280 mmHg – Dog/Horse
Measurement Time: 30 to 50 seconds typical, 120 seconds maximum
Default Inflation Pressure: 200 mmHg – Cat
150 mmHg – Dog
120 mmHg – Horse
Calibration:
Factory calibrated
AUTO Interval Times:
1, 2, 3, 4, 5, 10, 15, 20, or 30 minutes
1, 2, 4, or 8 hours
Alarm Range:
20-260 (in 1 mmHg steps), and OFF
Pulse Rate Accuracy:*
25- 250 bpm, ± 2% or ± 3 bpm (whichever is greater)
251-300 bpm, ± 5%
*NOTE: NIBP Pulse Rate accuracy outside of this range is unspecified.
Altitude Range:
3048m (10,000 ft)
Power Requirements
AC Input:
100 to 240 V, 0.8 A, 50/60 Hz
Dimensions
Width:
18.0 cm (7 inches)
Length:
34.0 cm (13.25 inches)
Height:
30.5 cm (12.0 inches)
Weight:
4.1-5 kg (9-11 lbs.)
Environmental
Temperature:
0 to 40°C (32 to 104 °C) Operating
-40 to 75°C (-40 to 167 °F) Storage
Humidity:
15 to 95% (non-condensing) Operating
10 to 95% (non-condensing) Storage
NOTE! The monitor may not meet performance specifications when stored or used outside the
temperature and humidity ranges listed above.
Содержание SurgiVet Advisor WWV9230
Страница 2: ......
Страница 10: ...Table of Contents viii Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 68: ...Chapter 7 ECG 7 14 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 80: ...Chapter 8 Oximetry 8 12 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 106: ...Chapter 12 Service Menu 12 8 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 120: ...Appendix D Revision History D 2 Veterinary 3 Parameter Advisor Operation Manual This page is intentionally left blank ...
Страница 121: ......