Troubleshooting
69
40-5648-51B
• Utilities
to review alarm history, to review infu
-
sion history, to set time and date, to update peri-
odic maintenance timestamp.
• Update Firmware
to reprogram the pump’s
software version through the serial interface (only
available with service upgrade software diskette
and instructions). [
Not available on Medfusion™
3500 pumps version 4.1 and above
.]
• Set Language
to change the national language
displayed. [
Available on Medfusion™ 3500
pumps
.]
B
iomed
>
c
aliBration
Within
Biomed > Calibration
the following selec
-
tions are available. See the
Calibration
section of this
manual for further details.
1. CAL SIZE AND POSITION
5. CAL SYRINGE SIZE SENSOR
2. CAL FORCE SENSOR
6. ADJUST CONTRAST
3. CAL PRESSURE SENSOR
7. VIEW CALIBRATION DATA
4. CAL PLUNGER POSTITION
8. SAVE CHANGES AND EXIT
PRESS THE NUMBER TO SELECT THE MODE
PRESS THE NUMBER TO SELECT
1
CAL SIZE AND POSITION
5
ADJUST CONTRAST
2
CAL FORCE SENSOR
6
VIEW CALIBRATION DATA
3
CAL PLUNGER POSITION
7
SAVE CHANGES AND EXIT
4
CAL SYRINGE SIZE SENSOR
B
iomed
>
d
iagnostics
(screen 1 of 2)
Within
Biomed > Diagnostics
the following selec
-
tions are available (one of two screens). Use the
num-
ber
buttons to select the test.
1. AUDIO TEST
5. MONITOR ANALOG SENSORS
2. DISPLAY TEST
6. MONITOR DIGITAL SENSORS
3. INDICATOR TEST
7. MONITOR BATTERY STATUS
4. KEYPAD TEST
PRESS THE NUMBER TO SELECT THE MODE
MORE
PRESS THE NUMBER TO SELECT
MORE
1
AUDIO TEST
5
MONITOR ANALOG SENSORS
2
DISPLAY TEST
6
MONITOR DIGITAL SENSORS
3
INDICATOR TEST
7
MONITOR BATTERY STATUS
4
KEYPAD TEST
8
DRIVE TRAIN TEST
a
udio
t
est
Use the
number
buttons to select the test.
Set Alarm
Style
allows testing of the various alarm tones.
Test
Speaker
is a factory/engineering test.
1. SET ALARM STYLE
2. TEST SPEAKER
PRESS THE NUMBER TO SELECT THE MODE
PRESS THE NUMBER TO SELECT
1
SET ALARM STYLE
2
TEST SPEAKER
d
isplay
t
est
Display Test
repeats blank, all pixels on, vertical
lines, horizontal lines, and checkerboard patterns
across the LCD display. This test may only be exited by
use of the Stop button.
STARTING THE DISPLAY TEST
PRESS STOP TO EXIT
STARTING THE DISPLAY TEST
PRESS
<
STOP
>
TO EXIT
i
ndicator
t
est
The
Indicator Test
allows the individual control
and testing of each indicator on the front panel (ex
-
cept battery and AC). Use the
number
button for the
indicator to toggle on or off.
1. RED ALARM OFF
5. MID INFUSE OFF
2. YELLOW ALARM OFF
6. RIGHT INFUSE OFF
3. PRESSURE CELL OFF
7. LOCK OFF
4. LEFT INFUSE OFF
PRESS THE NUMBER TO CHANGE LED STATE
PRESS THE NUMBER TO CHANGE LED STATE
1
RED ALARM OFF
5
MID INFUSE OFF
2
YELLOW ALARM OFF
6
RIGHT INFUSE OFF
3
PRESSURE CELL OFF
7
LOCK OFF
4
LEFT INFUSE OFF
Содержание Medfusion 3000 Series
Страница 1: ...s Medfusion 3000 Series Technical Service Manual ...
Страница 2: ...Table of Contents ii 40 5648 51B ...
Страница 20: ...Introduction 14 40 5648 51B ...
Страница 32: ...Scheduled Maintenance 26 40 5648 51B ...
Страница 80: ...Troubleshooting 74 40 5648 51B ...
Страница 116: ...Calibration Adjustment 110 40 5648 51B ...
Страница 158: ...Assembly Drawings Parts Lists 152 40 5648 51B Medfusion 3000 series main pump assembly See Note 1 ...
Страница 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Страница 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Страница 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Страница 167: ...Assembly Drawings Parts Lists 161 40 5648 51B Barrel clamp assembly GREASE SILICONE III SOLDER NO CLEAN ...
Страница 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Страница 182: ...Index 176 40 5648 51B ...
Страница 183: ......