Scheduled Maintenance
17
40-5648-51B
Periodic maintenance
This maintenance is
required
for the continued safe
operation of Medfusion™ 3000 Series pumps.
The Medfusion™ 3000 Series pump must be tested
annually
, or whenever the pump has been damaged
or dropped. Always check all sensor calibrations as a
standard part of annual maintenance. No calibration is
required to maintain pump flow delivery accuracy.
Installation/quick check-out
Each pump is inspected and tested prior to shipment
from the factory. Some institutions require that devices
entering a hospital be functionally checked before being
placed into service. The following procedure is provided
to meet this need.
Quick check-out test
c
1) Plug AC power cord into AC receptacle on side of pump, then plug pump into an AC power (100 to
240 VAC) source. Verify AC indicator and battery charge indicator are lit.
c
2) Press
Power
. Verify the following on power-up:
• Verify alarm beeps.
• Verify all the LED indicators are turned on & off (except AC and battery).
• Verify that no self-tests fail.
• Verify the display is legible and contrast is acceptable.
c
3) Unplug the AC power cord and verify that within several seconds the
battery indicator begins to
flash and AC indicator goes off. Re-plug the AC power cord into the unit.
c
4) Perform the “Flow Delivery Accuracy Test” from
Annual Maintenance
(which follows).
c
5) From the pump’s
Main
screen, use the number buttons to choose
Biomed
(press
More
for second
page, if needed). Use the number buttons to enter passcode “
2580
”, then use the number buttons to
choose
Utilities
. Then:
Set/view Last PM Date
.
Pumps with V3 software:
Set the PM date to today’s date. (Setting this
PM date means that an advisory message will appear on the pump two years from this date.)
- OR -
Set/view Next PM Date
.
Pumps with V4 software:
Set the date to the date you
want
the PM alert
to occur.
This message does not prevent normal operation of the pump, but advises the user that maintenance
is recommended.) [Note: pumps are shipped from the factory with the date/time set to USA Central
Standard Time; use the
Set Time/Date
utility to modify this if desired.]
c
6) Perform the “AC Line Leakage Test” from
Annual Maintenance
(which follows).
c
7) Consult the troubleshooting guide or contact Smiths Medical should the pump fail any steps in this
test.
c
8) After the completion of these tests the pump be should plugged into AC power to recharge the bat
-
tery. [This is recommended when the battery capacity is less than 90% – look at battery gauge dis
-
played on the bottom of the power-up screen.]
Содержание Medfusion 3000 Series
Страница 1: ...s Medfusion 3000 Series Technical Service Manual ...
Страница 2: ...Table of Contents ii 40 5648 51B ...
Страница 20: ...Introduction 14 40 5648 51B ...
Страница 32: ...Scheduled Maintenance 26 40 5648 51B ...
Страница 80: ...Troubleshooting 74 40 5648 51B ...
Страница 116: ...Calibration Adjustment 110 40 5648 51B ...
Страница 158: ...Assembly Drawings Parts Lists 152 40 5648 51B Medfusion 3000 series main pump assembly See Note 1 ...
Страница 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Страница 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Страница 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Страница 167: ...Assembly Drawings Parts Lists 161 40 5648 51B Barrel clamp assembly GREASE SILICONE III SOLDER NO CLEAN ...
Страница 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Страница 182: ...Index 176 40 5648 51B ...
Страница 183: ......