Scheduled Maintenance
25
40-5648-51B
to a level which disconnects the battery gauge.
The
pump then loses all its calibration information
.
From the time the
Battery Depleted
alarm
occurs, the pump has approximately 5 minutes of
operation on battery before the “fully-depleted dis-
connect” condition occurs. At this point, the audible
alarm continues running from the super-capacitor
on the main board.
6.
After several minutes of recharging the gauge will
show the learned capacity of the battery.
7.
Turn the pump on and select
Biomed
>
Diag-
nostics
>
Monitor Battery Status
.
8.
Verify LMD (last measured discharge) value is
greater than 1600 ma-hours.
9.
Press
More
and verify CPI is 0×00 or 0×01.
10.
Turn the pump off and continue to charge pump
until battery charge is complete (approximately 10
hours).
11.
Turn on pump and select
Biomed
>
Diagnos-
tics
>
Monitor Battery Status
.
12.
Verify the LMD is greater than
1600 mA-hours
. At
1600 mA-hours battery, the pump’s battery life is
approximately 7-8 hours. It is recommended that
the battery be replaced if its capacity is less than
1600 mA-hours.
Requirements for battery pack
replacement
The battery must be replaced only by a trained Bio-
medical or service technician. Always dispose of
depleted or defective batteries in compliance with all
applicable regulations or you may return the battery
pack to Smiths Medical for recycling.
WARNING:
Battery Replacement:
Observe ESD
handling precautions when replacing the battery.
Replace battery only with same Smiths Medical
part/model number.
Recycle batteries in compliance with applicable local
regulations.
Collect Separately
This product contains electrical and electronic com-
ponents (including batteries) that may contain materi-
als, which if disposed of with general waste, could be
damaging to the environment.
In accordance with Directive 2002/96/EC Waste
Electrical and Electronic Equipment, residents of the
European Union must follow specific disposal or recy
-
cling instructions for this product. Contact your local
distributor, or visit the following web site for specific
instructions:
http://www.smiths-medical.com/recycle/index.html
Non-European Union residents must dispose of or re
-
cycle this product (including batteries) in accordance
with the local laws or regulations that apply.
WARNING:
Collect Separately
. There are potential
health hazards associated with improper disposal
of batteries, electronics, and contaminated (used)
infusion sets and syringes. Dispose of used batteries,
infusion sets, syringes, and other used accessories,
or a pump that has reached the end of its useful life,
in an environmentally safe manner, and according
to any regulations that may apply.
Содержание Medfusion 3000 Series
Страница 1: ...s Medfusion 3000 Series Technical Service Manual ...
Страница 2: ...Table of Contents ii 40 5648 51B ...
Страница 20: ...Introduction 14 40 5648 51B ...
Страница 32: ...Scheduled Maintenance 26 40 5648 51B ...
Страница 80: ...Troubleshooting 74 40 5648 51B ...
Страница 116: ...Calibration Adjustment 110 40 5648 51B ...
Страница 158: ...Assembly Drawings Parts Lists 152 40 5648 51B Medfusion 3000 series main pump assembly See Note 1 ...
Страница 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Страница 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Страница 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Страница 167: ...Assembly Drawings Parts Lists 161 40 5648 51B Barrel clamp assembly GREASE SILICONE III SOLDER NO CLEAN ...
Страница 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Страница 182: ...Index 176 40 5648 51B ...
Страница 183: ......