Parts Replacement
76
40-5648-51B
Opening & closing the
pump housing
Tools needed
For this disassembly procedure you will need:
• Torque screwdriver with a #1 Phillips head
Always work at an electrostatic-controlled work sta-
tion
when disassembling the pump.
Opening the pump housing
1. Unplug the AC power cord (also disconnect any
external DC power, if in use).
2. If the poleclamp is attached to the pump, you must
remove it before disassembling the pump. Remove
the 2 flat head screws attaching the poleclamp
bracket to bottom housing of the pump.
3. Remove the 2 flat head screws from the
battery
compartment cover, and then remove the cover.
4. Carefully remove battery pack from its compart
-
ment, and disconnect its cable from the intercon
-
nect board.
5. Remove the 2 pan head screws from the bottom
housing of the pump.
6. Carefully separate the two halves of the pump
housing (there are three snap tabs across the
back).
7. Unplug the ribbon cable from J9 on the main
board (in most cases the other end of this ribbon
cable is glued or soldered into the interconnect
board).
Closing the pump housing
1. Before closing the pump housing, be sure the AC
and external DC power are disconnected, and the
battery pack is removed.
2. Reconnect the ribbon cable joining the intercon
-
nect board and the main board at J9.
3. Be sure all wires are clear of pinch points when
rejoining the housing halves.
4. There are 3 snap fit tabs and slots along the back
-
side of the housing. Align them and close the top
& bottom housings together.
5. Screw in the 2 pan head screws which connect the
top & bottom housings. Torque these screws to 60
in oz (0.8 Nm).
6. Orient the battery pack with the circuit board and
cable facing the rectangular cutout in the battery
compartment.
7. Connect battery cable to the interconnect board.
Slide the battery into the compartment.
8. Replace battery cover and secure it with 2 flathead
screws. Torque these screws to 60 in oz (0.8 Nm).
9. If using the pole clamp, reattach it to the pump
with the 2 flathead screws. Torque these screws to
160 in oz (2.3 Nm).
Follow retest guidelines listed in the Calibration and
Adjustment section to verify pump functionality be
-
fore returning it to use.
Содержание Medfusion 3000 Series
Страница 1: ...s Medfusion 3000 Series Technical Service Manual ...
Страница 2: ...Table of Contents ii 40 5648 51B ...
Страница 20: ...Introduction 14 40 5648 51B ...
Страница 32: ...Scheduled Maintenance 26 40 5648 51B ...
Страница 80: ...Troubleshooting 74 40 5648 51B ...
Страница 116: ...Calibration Adjustment 110 40 5648 51B ...
Страница 158: ...Assembly Drawings Parts Lists 152 40 5648 51B Medfusion 3000 series main pump assembly See Note 1 ...
Страница 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Страница 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Страница 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Страница 167: ...Assembly Drawings Parts Lists 161 40 5648 51B Barrel clamp assembly GREASE SILICONE III SOLDER NO CLEAN ...
Страница 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Страница 182: ...Index 176 40 5648 51B ...
Страница 183: ......