Scheduled Maintenance
19
40-5648-51B
Mandatory annual maintenance testing
All tests on this
Required Annual Maintenance
List
must
be performed annually in order to ensure the continued
safe operation of the Medfusion™ 3000 Series pump.
General inspection
c
1) If not already performed, clean and disinfect the pump as described in the “
Cleaning and Care
” sec-
tion (see previous page).
c
2) Inspect for obvious physical damage, including cracked housings or torn
keypads. Repair any physi-
cal damage.
c
3) Verify smooth operation of syringe plunger driver, syringe release lever, syringe flange clip, and sy
-
ringe barrel clamp. Clean and/or repair any damaged components.
c
4) Verify the three tubing guides (hooks) on top left side are intact, and fully secured to pump housing.
Replace any damaged guides.
c
5) Verify force sensor seal, located behind flippers on plunger driver head, is intact and not punctured.
Replace damaged seal.
c
6) Inspect power cord for damage and wear. Replace damaged or worn power cord before performing
any leakage current testing.
c
7) Inspect pole clamp for proper operation. Verify that the screws holding pole clamp components are
tight.
c
8) Verify the Serial Number label is legible and intact. Replace if label is damaged.
c
9) Verify the Warning label is legible and intact. Replace if label is damaged.
c
10) Verify four rubber feet are attached to the bottom housing of the pump. Replace missing feet.
c
11) Turn the pump on all sides and check for any loose parts internally and externally.
c
12) Plug AC power cord into AC receptacle on side of pump, then plug pump into an AC power (100 to
240 VAC) source. Verify AC indicator and battery charge indicator are lit. Consult troubleshooting
for any failed indicator.
Power-up test
Ensure that no syringe is loaded in the pump and the
barrel clamp is fully down.
Press
ë
. Verify the follow
-
ing on power-up:
c
1) Verify alarm beeps.
c
2) Verify all the LED indicators are turn on and off (except AC and battery).
c
3) Verify that battery capacity shown on screen is greater than 0%.
c
4) Verify that no self-tests fail.
c
5) Verify the display is legible and contrast is acceptable.
c
6) Unplug the AC power cord and verify that within several seconds the battery indicator begins to
flash and AC indicator goes off. Re-plug the power cord into the unit.
Consult troubleshooting for any failed steps in this test.
Содержание Medfusion 3000 Series
Страница 1: ...s Medfusion 3000 Series Technical Service Manual ...
Страница 2: ...Table of Contents ii 40 5648 51B ...
Страница 20: ...Introduction 14 40 5648 51B ...
Страница 32: ...Scheduled Maintenance 26 40 5648 51B ...
Страница 80: ...Troubleshooting 74 40 5648 51B ...
Страница 116: ...Calibration Adjustment 110 40 5648 51B ...
Страница 158: ...Assembly Drawings Parts Lists 152 40 5648 51B Medfusion 3000 series main pump assembly See Note 1 ...
Страница 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Страница 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Страница 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Страница 167: ...Assembly Drawings Parts Lists 161 40 5648 51B Barrel clamp assembly GREASE SILICONE III SOLDER NO CLEAN ...
Страница 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Страница 182: ...Index 176 40 5648 51B ...
Страница 183: ......