Calibration & Adjustment
109
40-5648-51B
Replace case top
Recalibrate all sensors
Complete Mandatory Annual Mainte-
nance Testing
Replace case handle
Recalibrate position & size
Quick Check-out & Calibration Verifi
-
cation
Replace plunger head,
or flippers
Recalibrate force
Quick Check-out & Calibration Verifi
-
cation
Replace IR lens
None
Quick Check-out
Replace plunger lever
None
Quick Check-out
Replace pump labels
None
Quick Check-out
Replace feet
None
Quick Check-out
Replace tubing holders
Recalibrate size
Quick Check-out & Calibration Verifi
-
cation
Содержание Medfusion 3000 Series
Страница 1: ...s Medfusion 3000 Series Technical Service Manual ...
Страница 2: ...Table of Contents ii 40 5648 51B ...
Страница 20: ...Introduction 14 40 5648 51B ...
Страница 32: ...Scheduled Maintenance 26 40 5648 51B ...
Страница 80: ...Troubleshooting 74 40 5648 51B ...
Страница 116: ...Calibration Adjustment 110 40 5648 51B ...
Страница 158: ...Assembly Drawings Parts Lists 152 40 5648 51B Medfusion 3000 series main pump assembly See Note 1 ...
Страница 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Страница 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Страница 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Страница 167: ...Assembly Drawings Parts Lists 161 40 5648 51B Barrel clamp assembly GREASE SILICONE III SOLDER NO CLEAN ...
Страница 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Страница 182: ...Index 176 40 5648 51B ...
Страница 183: ......