Troubleshooting
55
40-5648-51B
Alarm message
Alarm
priority
programming
infusing
Cause or corrective action
C
heCk
S
yringe
P
lunger
S
enSor
High
–
At power-up, software has sensed the plunger is
loaded when no syringe is loaded in the pump
(as sensed by the barrel clamp size sensor). Use
Biomed
>
Diagnostics
and run
Monitor Digi-
tal Sensors
to check function of both plunger
sensors.
C
lear
tvd
to
S
tart
i
nfuSion
Limit
–
Normal advisory message. Total volume delivered
(TVD) or total dose delivered (TDD) needs to be
cleared to start an infusion when TVD/TDD ex
-
ceeds the volume limit setting.
C
omPleting
S
yStem
i
nitialization
Limit
–
Normal advisory message. At power-up, if all set
-
tings are entered by the user and
É
is pressed
before background self-tests are completed, then
this message appears. Wait a few seconds until self-
testing is completed and then press
É
to begin
infusion.
d
ata
d
oeS
n
ot
m
atCh
P
reviouS
e
ntry
Limit
–
Normal advisory message. In cases where confirma
-
tion of user entered data is required, the confirma
-
tion value did not match. Re-enter the value.
d
ePleted
B
attery
(S
yStem
f
ailure
)
High
High
Normal alarm message. Battery voltage measured
by the gauge on the battery pack is too low. Software
then stops delivery and allows watchdog to generate
alarm. Plug pump into AC power, cycle power on
the pump, then use
Recall Last Settings
to re-
call delivery settings and resume delivery. If problem
is chronic, check battery pack function or perform
battery calibration.
f
orCe
S
enSor
B
gnd
t
eSt
High
High
Normal alarm message. Press
ë
to turn the pump
off. Background self-test has found the force sensor
signal out-of-range or the force sensor signal did not
change when bridge-test signal was asserted. Use
Biomed > Diagnostics > Monitor Analog
Sensors
to check operation of plunger force sen
-
sor. Open plunger head and check plunger cable and
force sensor cable on the plunger board. Replace
plunger board, plunger cable or force sensor.
i
nfuSion
C
omPlete
–
High
Normal alarm message. Infusion has reached the set
volume limit and delivery has stopped.
i
nPut
o
ut
of
r
ange
-
g
reater
t
han
m
ax
v
alue
Limit
–
Normal advisory message. User attempted to enter
a number greater than the maximum value allowed.
(See also “maximum rate” in general problems be-
low). Try a lower value.
Содержание Medfusion 3000 Series
Страница 1: ...s Medfusion 3000 Series Technical Service Manual ...
Страница 2: ...Table of Contents ii 40 5648 51B ...
Страница 20: ...Introduction 14 40 5648 51B ...
Страница 32: ...Scheduled Maintenance 26 40 5648 51B ...
Страница 80: ...Troubleshooting 74 40 5648 51B ...
Страница 116: ...Calibration Adjustment 110 40 5648 51B ...
Страница 158: ...Assembly Drawings Parts Lists 152 40 5648 51B Medfusion 3000 series main pump assembly See Note 1 ...
Страница 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Страница 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Страница 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Страница 167: ...Assembly Drawings Parts Lists 161 40 5648 51B Barrel clamp assembly GREASE SILICONE III SOLDER NO CLEAN ...
Страница 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Страница 182: ...Index 176 40 5648 51B ...
Страница 183: ......