Theory of Operation
47
40-5648-51B
Speaker test
The
return signal from the speaker detects whether
current is flowing in the Speaker from analog condi
-
tioning circuitry. The half-wave rectifier circuit uses
the
negative half-cycle
of the sense signal to invert and
amplify it.
•
The amplified output is demodulated by the analog
switch which creates a peak and hold circuit.
•
The
speaker_detect output signal from the
conditioning circuitry goes to the microprocessor’s
analog input.
Plunger board – schematic level
The plunger PCB provides pre-amplification of the
force sensor output to the Main system PCB, and
contains two photo-interrupters with supporting cir-
cuitry, one for each plunger flipper.
The referenced schematic signals of the main board
electronics are identified using names in
Smallcaps
.
Force preamplifier function
The force sensing system provides a measurement of
the force required in moving the syringe plunger. The
interface to the force sensor has an excitation source,
test signal, and pre-amplification.
•
The Main printed circuit board provides a fixed
4.096-volt excitation to the force sensor.
•
The Main printed circuit board trims this offset
range using the signal
force_preamp_osadj
.
•
The Main PCB provides additional gain to maxi
-
mize measurement resolution.
Force sensing interface description
The force sensing element is a full bridge strain gage,
and the plunger PCB pre-amplifies and filters the out-
put of this bridge. The interface has:
sensor excitation,
electronic zero-offset adjustment, amplification and
filtering, and a sensor test control signal
•
force_snsr_exh provides a constant voltage
excitation of
4.096 volts DC
to the sensor.
•
Auto-Zero Adjustment Circuit provides a voltage
to zero the pre-amplifier,
force_osv_adj, from
the 8 bit system D/A converter.
•
Sensor Post Filter Amplifier is a second order low-
pass filter with a 3.68x gain.
•
The signal
force_snsr_test* when set active low
causes Q2 to apply a resistance parallel to the force
sensor’s bridge resulting in a positive shift of the
sensor output.
•
The sensing system requires calibration to de-
termine zero adjustment, offset, and gain. Cali-
bration values are stored in non-volatile (serial
EEPROM) memory.
Содержание Medfusion 3000 Series
Страница 1: ...s Medfusion 3000 Series Technical Service Manual ...
Страница 2: ...Table of Contents ii 40 5648 51B ...
Страница 20: ...Introduction 14 40 5648 51B ...
Страница 32: ...Scheduled Maintenance 26 40 5648 51B ...
Страница 80: ...Troubleshooting 74 40 5648 51B ...
Страница 116: ...Calibration Adjustment 110 40 5648 51B ...
Страница 158: ...Assembly Drawings Parts Lists 152 40 5648 51B Medfusion 3000 series main pump assembly See Note 1 ...
Страница 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Страница 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Страница 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Страница 167: ...Assembly Drawings Parts Lists 161 40 5648 51B Barrel clamp assembly GREASE SILICONE III SOLDER NO CLEAN ...
Страница 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Страница 182: ...Index 176 40 5648 51B ...
Страница 183: ......