Troubleshooting
60
40-5648-51B
System Advisory Alarms
This section explains the system advisory alarms in the
Medfusion™ 3000 Series pump, and suggests possible
causes and remedies for these alarms.
System advisory alarm
Cause or remedy
B
aCkuP
C
ritiCal
d
ata
C
orruPted
Configuration and library settings are stored redundantly in non-
volatile serial EEPROM on the main board. This message will occur
if backup data was lost, and primary data was used to restore back
-
up values. No action is required.
d
efault
C
onfiguration
r
eStored
Configuration settings are stored redundantly in non-volatile serial
EEPROM on the main board. If both primary and backup are lost,
the configuration is defaulted. Use pump
Custom Program
>
Teach/Learn
to copy configuration from a known good pump to
restore settings. If problem persists, there may be a problem with
the main board. This message may appear when a new main board
is installed.
h
iStory
d
ata
C
orruPted
Infusion history log is stored in non-volatile serial EEPROM on the
main board. This message will occur if this data is lost. No action is
required.
i
mProPer
S
hutdoWn
This alarm is recorded to history when the pump is turned on if
the last power-down did not occur through pressing the power key.
Check for possible causes for a pump power failure.
i
nvalid
i
nterruPt
& S
yStem
a
dviSory
:
i
mProPer
S
hutdoWn
Replace the Main PCB.
n
on
C
ritiCal
d
ata
C
orruPted
Other settings are stored in non-volatile serial EEPROM on the
main board. This message will occur if this data was lost. No action
is required.
P
harm
g
uard
d
ata
C
orruPted
The PharmGuard
®
Safety Data is stored in non-volatile serial EE
-
PROM on the main board. This message will ocur if this data is lost.
No action is required.
P
rimary
C
ritiCal
d
ata
C
orruPted
Configuration and library settings are stored redundantly in non-
volatile serial EEPROM on the main board. This message will occur
if primary data was lost, and backup data was used to restore pri
-
mary values. No action is required.
S
et
t
ime
and
d
ate
Advisory message that real time
clock has reset on the pump. This
may occur if the battery is depleted and the main board supply
totally discharged – the super-capacitor on main board should keep
up the clock for many days after a depleted battery is reached. Use
Biomed
>
Utilities
and run
Set Time/Date
to restore date and
time settings. No functionality of pump is affected by date/time set
-
ting.
m
aintenanCe
i
S
r
eCommended
This is a reminder to the user to conduct the Annual Maintenance
Testing (section 1, page 19). To clear this alarm, reset the PM Main
-
tenance date (see section 5, Quick Maintenance Check-out Test and
follow the instructions).
Содержание Medfusion 3000 Series
Страница 1: ...s Medfusion 3000 Series Technical Service Manual ...
Страница 2: ...Table of Contents ii 40 5648 51B ...
Страница 20: ...Introduction 14 40 5648 51B ...
Страница 32: ...Scheduled Maintenance 26 40 5648 51B ...
Страница 80: ...Troubleshooting 74 40 5648 51B ...
Страница 116: ...Calibration Adjustment 110 40 5648 51B ...
Страница 158: ...Assembly Drawings Parts Lists 152 40 5648 51B Medfusion 3000 series main pump assembly See Note 1 ...
Страница 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Страница 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Страница 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Страница 167: ...Assembly Drawings Parts Lists 161 40 5648 51B Barrel clamp assembly GREASE SILICONE III SOLDER NO CLEAN ...
Страница 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Страница 182: ...Index 176 40 5648 51B ...
Страница 183: ......