Troubleshooting
54
40-5648-51B
Alarm messages & priorities
The following table defines alarm message, alarm prior
-
ity, and cause for each alarm condition:
Alarm message
Alarm
priority
programming
infusing
Cause or corrective action
B
attery
C
ommuniCation
t
imeout
(S
yStem
a
dviSory
)
Low
Low
Pump has sensed battery current present, but the
battery gauge on the battery pack is not commu-
nicating. Check connections from main board, to
interconnect, to battery pack.
B
attery
n
ot
C
harging
(S
yStem
a
dviSory
)
Low
Low
Pump has sensed AC or DC external power is pres
-
ent, but the battery is not charging. Use
Biomed
>
Diagnostics
and run
Monitor Battery Sta-
tus
to review battery function. Check battery con
-
nections and battery charge circuit on the intercon-
nect board.
B
attery
n
ot
W
orking
(S
yStem
a
dviSory
)
Medium
Medium
Pump has found no battery present – no charge
current and no battery gage communication. Battery
may be unplugged.
C
alCulated
r
ate
o
ut
of
r
ange
High
–
The calculated rate is not valid. Verify that all infu
-
sion parameters have been entered correctly.
C
heCk
C
lutCh
/ P
lunger
l
ever
–
High
Normal alarm message. Software uses position po
-
tentiometer to monitor motion of the plunger driver
during delivery. If the motion is not correct this
alarm is generated.
May be caused because pump was not primed and
clutches took too long to engage, or may occur when
user squeezes the plunger lever and moves plunger
head.
May also occur if position pot is not calibrated or is
contaminated – use
Biomed
>
Utilities
and run
View Alarm History
to check if failure occurs at
same position reading.
C
heCk
S
yringe
B
arrel
C
lamP
Medium
–
Certain syringe brands and sizes may cause the bar-
rel clamp sensor to have difficulty in verifying the
clamp is in position. Use
Biomed
>
Diagnostics
and run
Monitor Digital Sensors
to check
function of barrel sensor.
C
heCk
S
yringe
f
lange
S
enSor
High
–
At power-up, software has sensed the flange is load
-
ed when no syringe is loaded in the pump (as sensed
by the barrel clamp size sensor). Use
Biomed
>
Di-
agnostics
and run
Monitor Digital Sensors
to check function of flange sensor.
Содержание Medfusion 3000 Series
Страница 1: ...s Medfusion 3000 Series Technical Service Manual ...
Страница 2: ...Table of Contents ii 40 5648 51B ...
Страница 20: ...Introduction 14 40 5648 51B ...
Страница 32: ...Scheduled Maintenance 26 40 5648 51B ...
Страница 80: ...Troubleshooting 74 40 5648 51B ...
Страница 116: ...Calibration Adjustment 110 40 5648 51B ...
Страница 158: ...Assembly Drawings Parts Lists 152 40 5648 51B Medfusion 3000 series main pump assembly See Note 1 ...
Страница 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Страница 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Страница 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Страница 167: ...Assembly Drawings Parts Lists 161 40 5648 51B Barrel clamp assembly GREASE SILICONE III SOLDER NO CLEAN ...
Страница 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Страница 182: ...Index 176 40 5648 51B ...
Страница 183: ......