Troubleshooting
53
40-5648-51B
Problem solving alarms / alerts
This section defines basic problems and provides some
standard problem solving procedures.
Types of alarms / alerts
The following table defines the alarms and alerts gener
-
ated by the Medfusion™ 3000 Series pumps, with sug
-
gested remedies and solutions.
Type
Definition and Remedy
High priority
A
high priority alarm
results from any condition which halts an ongoing infusion,
or any pump system fault which affects infusion. High Priority alarms are signaled
by a flashing red indicator and an audible signal.
Press
í
to pause the audible alarm for the preset alarm silence period.
Medium priority
A
medium priority alarm
indicates any condition requiring operator intervention
but does not halt infusion. Medium Priority alarms are signaled with a flashing yel
-
low indicator and an audible signal.
Pressing
í
silences the audible alarm for the programmed alarm silence period.
Low priority
A
low priority alarm
indicates any condition not requiring immediate operator
intervention. Low Priority alarms are signaled with a continuous yellow indicator
and an intermittent audible signal.
Pressing
í
permanently silences this alarm.
Limit priority
A
limit alarm
occurs whenever an invalid entry is attempted on a numerical en
-
try screen. The invalid entry alarm sounds a brief tone with an advisory message
onscreen.
Neglected pump
The
Neglected Pump alarm
is a medium priority alarm. It is simply reminding you
to finish what you started. Once you begin programming any infusion delivery set
-
ting, the pump expects you to continue until setup is complete.
If you leave the pump paused too long (>30 seconds) on a data entry screen, then
the pump sounds a medium priority alarm tone with the yellow indicator flashing.
Note:
The Neglected Pump alarm, the Periodic Callback
alarm, the Improper Shutdown alarm, the Check
Syringe Barrel Clamp alarm, and all Limit Priority
alarms are not recorded in the pump’s alarm history
record.
Содержание Medfusion 3000 Series
Страница 1: ...s Medfusion 3000 Series Technical Service Manual ...
Страница 2: ...Table of Contents ii 40 5648 51B ...
Страница 20: ...Introduction 14 40 5648 51B ...
Страница 32: ...Scheduled Maintenance 26 40 5648 51B ...
Страница 80: ...Troubleshooting 74 40 5648 51B ...
Страница 116: ...Calibration Adjustment 110 40 5648 51B ...
Страница 158: ...Assembly Drawings Parts Lists 152 40 5648 51B Medfusion 3000 series main pump assembly See Note 1 ...
Страница 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Страница 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Страница 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Страница 167: ...Assembly Drawings Parts Lists 161 40 5648 51B Barrel clamp assembly GREASE SILICONE III SOLDER NO CLEAN ...
Страница 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Страница 182: ...Index 176 40 5648 51B ...
Страница 183: ......