Introduction
11
40-5648-51B
External Power Detector
Circuitry which determines when and if external power (whether AC or DC) has
been or is connected to the pump.
Flash Memory
The pump uses flash memory, which is re-programmable through the infrared (IR)
port and a remote computer.
High Priority Alarms
A high priority alarm results from either any condition which halts an ongoing in
-
fusion, or any pump system fault which affects infusion. If the front panel controls
are locked when a high priority alarm occurs,
the pump controls unlock
.
High Priority alarms are signaled by a flashing red indicator and an audible signal.
Press
í
to pause the audible alarm for the preset alarm silence period.
Infrared Serial Data Port
The infrared serial port interfaces directly with the main microprocessor’s asyn
-
chronous serial communication pins. The infrared port supports short transmission
distances of 2” or less and a maximum baud rate of 9600.
Infusion Class 4
An infusion pump which combines the functions of continuous infusion flow, in
-
termittent flow, and discrete bolus delivery.
Infusion Class 5
An infusion pump which functions as a profile pump, providing a programmed
sequence of delivery rates.
Interconnect Board
The interconnect printed circuit board interfaces to an intelligent NiMH
Battery
(or NiCad on older model pumps), the system speaker, the internal DC supply and
supply conditioning for an external DC supply and the main PCB.
IPX3
Equipment which is ingress protected from fluid spraying at a vertical angle from
above, and from angles to 60° on either side of vertical.
Keypad Test
Verifies individual function of each key on the keypad. Nonfunctioning keys indi
-
cate need for keypad replacement.
LCD
Liquid Crystal Display.
LCD Backlight
The LED fiber optic light source which illuminates the LCD display.
Limit Priority Alarms
A limit priority alarm is generated whenever a preset minimum or maximum limit
has been violated. For example: when programming an infusion there are minimum
and maximum limits preset rates assigned to syringes by size & manufacturer.
• The limit priority alarms sound a tone and display an advisory message on
-
screen for 3 seconds. To re-display the message press
í
.
LMD (Last Measured
Discharge)
This the learned capacity of the battery by the gauge following a calibration cycle.
Replace the battery when this is < 1600 mA-hours.
Low Priority Alarms
A low priority alarm indicates any condition not requiring immediate operator in-
tervention. Low Priority alarms are announced with a continuous yellow indicator
and an intermittent audible signal.
• Pressing
í
permanently silences this alarm.
• If the front panel controls are locked when a low priority alarm occurs,
the
pump controls do not unlock
.
Medium Priority Alarms
A medium priority alarm indicates any condition requiring operator intervention
but does not halt infusion. Medium Priority alarms are signaled with a flashing yel
-
low indicator and an audible signal. Pressing
í
will silence the audible alarm for
the programmed alarm silence period.
• If the front panel controls are locked when a medium priority alarm occurs,
the pump controls do not unlock
.
Содержание Medfusion 3000 Series
Страница 1: ...s Medfusion 3000 Series Technical Service Manual ...
Страница 2: ...Table of Contents ii 40 5648 51B ...
Страница 20: ...Introduction 14 40 5648 51B ...
Страница 32: ...Scheduled Maintenance 26 40 5648 51B ...
Страница 80: ...Troubleshooting 74 40 5648 51B ...
Страница 116: ...Calibration Adjustment 110 40 5648 51B ...
Страница 158: ...Assembly Drawings Parts Lists 152 40 5648 51B Medfusion 3000 series main pump assembly See Note 1 ...
Страница 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Страница 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Страница 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Страница 167: ...Assembly Drawings Parts Lists 161 40 5648 51B Barrel clamp assembly GREASE SILICONE III SOLDER NO CLEAN ...
Страница 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Страница 182: ...Index 176 40 5648 51B ...