Troubleshooting
65
40-5648-51B
General troubleshooting
This section discusses some potential problems which
may be encountered with the Medfusion™ 3000 Series
pump, and suggests remedies for those problems.
Problem
Remedy or solution
Alarms do not sound (either
loudly or at all)
a) If the alarms sound faintly, they may have been configured to their low
-
est level. Use
Custom Program
to change the audio alarm volume to a
higher level.
b) Use
Biomed > Diagnostics
and run the
Audio Test
on Speaker and
Alarm Style. Replace
speaker if audio alarm does not work correctly.
Problem:
No charge light
a) Connect to AC power. Turn on pump and allow to complete startup
self-testing. If both AC indicator and charge lights are off then see “Indica
-
tors Do Not Flash” below.
b) Use
Biomed > Diagnostics
and run the
Monitor Battery Status
test. Determine whether battery is charging at 50 MA or greater.
Battery Problem:
Battery not working message
a) Open the battery door in the case bottom and check the connection of
the battery ribbon cable to the interconnect board.
b) Use
Biomed > Diagnostics
and run the
Monitor Battery Status
test. If no information appears then replace the battery pack.
Battery Problem:
Battery does not hold charge
a) Open the battery door in the case bottom and check the connection of
the battery ribbon cable to the interconnect board.
b) Connect to AC power. Turn on pump and allow to complete startup
self-testing. Use
Biomed > Diagnostics
and run the
Monitor Bat-
tery Status
test. Check whether battery is charging at least +50 MA
(wait at least 1 minute from power-on for charge current reading to stabi
-
lize).
• Unplug AC power if pump immediately shuts down then replace bat
-
tery pack.
• If pump continues to operate and gauge shows normal discharge cur
-
rent -100 to -200 ma, then perform a battery calibration.
Indicator lights do not flash
If the indicator lights do not flash during startup, or along with their ap
-
propriate function or alarm, then:
• Open the case and check the keypad connection to the Main Board. If
any connections are loose, reconnect and re-test.
• Otherwise, replace keypad.
Keyclick (Beep) is faint or not
present
If the keyclicks (beeps) sound faintly, or not at all, they may have been
configured to their lowest level or turned off.
Use
Custom Program
to reset the audio to a higher level.
Keypad button does not work
Use
Biomed > Diagnostics
and run the
Keypad Test
. If any keys do
not pass this test, open the case and check connections for the keypad.
Liquid Crystal Display (LCD) has
poor contrast
Use
Biomed > Calibration
and run the
Adjust Contrast
.
If the LCD contrast cannot be adjusted, then check LCD negative supply,
and consider replacing display or main board.
Содержание Medfusion 3000 Series
Страница 1: ...s Medfusion 3000 Series Technical Service Manual ...
Страница 2: ...Table of Contents ii 40 5648 51B ...
Страница 20: ...Introduction 14 40 5648 51B ...
Страница 32: ...Scheduled Maintenance 26 40 5648 51B ...
Страница 80: ...Troubleshooting 74 40 5648 51B ...
Страница 116: ...Calibration Adjustment 110 40 5648 51B ...
Страница 158: ...Assembly Drawings Parts Lists 152 40 5648 51B Medfusion 3000 series main pump assembly See Note 1 ...
Страница 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Страница 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Страница 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Страница 167: ...Assembly Drawings Parts Lists 161 40 5648 51B Barrel clamp assembly GREASE SILICONE III SOLDER NO CLEAN ...
Страница 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Страница 182: ...Index 176 40 5648 51B ...
Страница 183: ......