Scheduled Maintenance
15
40-5648-51B
Section 1: Scheduled Maintenance
of the service man
-
ual defines the required preventive maintenance for
keeping the Medfusion™ 3000 Series syringe infusion
pump in good operating condition.
Preventive Maintenance
Planning
The recommended preventive maintenance plan al
-
lows you to service pumps in batches as their anniver
-
sary dates arrive. For example, if you have numerous
pumps to service over the course of a year you may
want to set the date on each pump so that 10 pumps
are due for service each week of the year (this will
prevent all pumps from giving the alert reminder at
the same time).
Pumps with
V3 software
: enter the
last
PM date (the
pump will calculate the next PM date). An example of
a plan for the maintenance of pumps with V3 software
is shown in the following table:
Serial
Numbers
Scheduled
PM
Enter Last
PM
Date of
Alarm
M001001 to
M001010
7/24/2007
7/24/2005
7/25/2007
M001011 to
M001020
7/31/2007
7/31/2005
8/1/2007
Pumps with
V4 software
: enter the
next
PM date. An
example of a plan for the maintenance of pumps with
V4 software is shown in the following table:
Serial
Numbers
Enter
Next PM
Date
Last PM
Date of
Alarm
M001001 to
M001010
7/24/2008
7/25/2007 7/25/2008
M001011 to
M001020
7/31/2008
8/1/2007
8/1/2008
The best approach is to set the date so that the alarm
occurs only if a pump is missed in the normal PM
schedule.
Biomed maintenance tools
In order to complete maintenance, parts replacement,
diagnosis and calibration of Medfusion™ 3000 Series
infusion pumps, you will need the following:
1 ea
−
Calibration Kit (see service parts list): Small
Calibration Slug, Large Calibration Slug & Force
Gauge
1 ea
−
new 50 or 60 cc syringe (note: the lubrication
in syringes evaporates, change “test” syringes
monthly)
1 ea
−
3-way Stopcock
1 ea
−
Electrical Safety Analyzer
1 ea
−
Torque Screwdriver w/Phillips & Standard bits.
1 ea
−
¼” open end wrench
1 ea
−
6” calipers with resolution of 0.001”
Using a torque screwdriver
You
must
always use a
Torque Screwdriver
when re-
assembling any Medfusion™ 3000 Series pump. Where
screws are used to secure components, over-tightening
can strip threads or crack standoffs in the case.
• Unless otherwise specified, you should always
torque all screws to 60
in-oz
(approximately 0.8
Nm).
Electrostatic-controlled
workstation
Whenever you work on the Medfusion™ 3000 Series
pump
−
specifically whenever you open the pump for
service
−
you
must
work in an electrostatic-controlled
environment. This ensures you will not damage the
electronic components in the pump.
WARNING: Repair Pump in ESD Controlled
Work Area:
The pump case should only be opened
at a work-station with Electrostatic controls, includ-
ing a grounded mat and wrist-strap.
Содержание Medfusion 3000 Series
Страница 1: ...s Medfusion 3000 Series Technical Service Manual ...
Страница 2: ...Table of Contents ii 40 5648 51B ...
Страница 20: ...Introduction 14 40 5648 51B ...
Страница 32: ...Scheduled Maintenance 26 40 5648 51B ...
Страница 80: ...Troubleshooting 74 40 5648 51B ...
Страница 116: ...Calibration Adjustment 110 40 5648 51B ...
Страница 158: ...Assembly Drawings Parts Lists 152 40 5648 51B Medfusion 3000 series main pump assembly See Note 1 ...
Страница 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Страница 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Страница 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Страница 167: ...Assembly Drawings Parts Lists 161 40 5648 51B Barrel clamp assembly GREASE SILICONE III SOLDER NO CLEAN ...
Страница 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Страница 182: ...Index 176 40 5648 51B ...
Страница 183: ......