4) A message dialog appears to confirm that you want to save the test result as a
reference.
Figure 72.
Confirmation dialog to use waveform as a reference.
5) If the results are good, click
Yes
. If the test needs to be run again, click
No
. Multiple
tests can be run to obtain the desired reference. In automatic mode, test results are
always saved. The reference will become part of the Test ID only when
Yes
is clicked
in the
Use as Reference
dialog box.
6) Motors that use the Test ID associated with the reference motor’s Test ID will now
have
Test-Ref EAR%
enabled in the Surge test setup window.
7) When you have finished the process, click
Save
then uncheck the
Edit Test ID
box to
turn off edit mode.
NOTE
If the reference waveform is no longer needed or if has been created by accident,
uncheck the Test-Ref EAR(%) and click the
Done
button, then update the Test ID and
it will detach the reference waveform from the selected Test ID.
68
PUB CM/I4 71-015 EN V13.2 Static Motor Analyzer—Baker AWA-IV User Manual
Baker AWA-IV software overview
Содержание Baker AWA-IV 12 HO
Страница 1: ...SKF Static Motor Analyzer Baker AWA IV User manual...
Страница 2: ......
Страница 22: ...2 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual About this manual...
Страница 126: ...106 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual Test procedures...
Страница 202: ...182 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual Database definition...
Страница 206: ...186 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual...
Страница 213: ......