Table 42.
Test ID table; temperature test parameters field descriptions.
Field name
Type
Description
temp_isAuto
Yes/No
1=Manual; 0=Automatic.
temp_degF_C
Yes/No
1=F degrees; 0=C degrees.
Table 43.
Test ID table; resistance test parameters field descriptions.
Field name
Type
Description
resist_no_leads
Short integer
# of leads (2 or 3) Used also in surge.
resist_deltaR_enabled
Yes/No
1=delta R used in pass/fail; 0=Not used.
resist_ deltaR_maxpercent
Short integer
Max difference in percent.
resist_external_enabled
Yes/No
Always false 0=No external testing.
resist_isAuto
Yes/No
1=Automatic; 0=Manual
resist_4lead_test_enabled
Yes/No
1=4 lead test enabled; 0=4 lead test disabled
(Only disabled at this time)
resist_correct_to_temp_enabled Yes/No
Correct resistance 1=Yes; 0=No.
resist_temp_correct_to
Short integer
Defines the value that the resistance
temperature will be corrected to.
resist_temp_correct_factor
Double
Default 234.5 for copper.
resist_target_coilRes_enabled
Yes/No
1=Coil resistance checking enabled; 0=not
enabled.
resist_target_coilRes
Double
Coil resistance to compare tested value with.
resist_target_coilRes_tolerance Long integer
Tolerance set for target coil resistance.
176
PUB CM/I4 71-015 EN V13.2 Static Motor Analyzer—Baker AWA-IV User Manual
Database definition
Содержание Baker AWA-IV 12 HO
Страница 1: ...SKF Static Motor Analyzer Baker AWA IV User manual...
Страница 2: ......
Страница 22: ...2 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual About this manual...
Страница 126: ...106 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual Test procedures...
Страница 202: ...182 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual Database definition...
Страница 206: ...186 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual...
Страница 213: ......