5) Click on the
HiPot/Step Voltage
dropdown list to select the
Step Voltage
test (for
this example). The
Ramp Voltage
test option is also selected from this list.
Figure 90.
DC Test parameters—Step Voltage parameters set.
6) When you select the
Step Voltage
option, the Config Wizard opens so you can
define the parameters needed for your test. The software will automatically
calculate values for the test based on other parameters entered, but you can adjust
the parameters as needed to fit your situation. When you have settled on the
parameters needed, close the Config Wizard to return to the
DC Test
setup window.
7) In the example above, we see that parameters are set for the Step Voltage test.
8) Click on the
Temp Correction
dropdown list to select the insulation type for you
motor. Options include
Thermoplastic
or
Thermosetting
. Choose None if you do
not want to set this parameter.
9) If you choose
Thermoplastic
or
Thermosetting
, the correction field will be enabled
so you can set the value needed for temperature correction.
10) When you are satisfied that your DC Tests have been configured as needed, click the
Close
button to return to the main test window.
88
PUB CM/I4 71-015 EN V13.2 Static Motor Analyzer—Baker AWA-IV User Manual
Test procedures
Содержание Baker AWA-IV 12 HO
Страница 1: ...SKF Static Motor Analyzer Baker AWA IV User manual...
Страница 2: ......
Страница 22: ...2 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual About this manual...
Страница 126: ...106 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual Test procedures...
Страница 202: ...182 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual Database definition...
Страница 206: ...186 PUB CM I4 71 015 EN V13 2 Static Motor Analyzer Baker AWA IV User Manual...
Страница 213: ......