9 Appendix
Installation Instructions Facescan
9.2 Conversion instructions for non-"Facescan ready" units
båÖäáëÜ
6. Unscrew the remaining screws from the protective plates (E).
7. Remove the covering plates (E) (top and bottom) from the connection
box (F) of board DX32.
8. The cable L3 can stay on the top covering plate (E).
9. Remove connector X2 from board DX32 and remove the protective
conductor.
10. Remove cable L2 from terminal X100 and pull it downwards from out
of the connection box (F).
11. Remove connector X1 from board DX32.
12. Loosen the two left-hand screws (G).
13. Loosen the two screws on the right (I) and remove the connection box
including board DX32.
9.2.3
Removing energy chain 1 (right)
Remove energy chain
L3
E
L3
G
DX32
X100
L2
X1
F
H
X2
I
IMPORTANT
Remove cable ties
For the following steps, all necessary cable ties should be removed with
wire cutters.